Abstract
The selection of an implant is a critical factor in the surgical treatment of patella fractures due to the risk of various complications, such as non-union, implant failure, and irritation. The present study evaluated and compared the biomechanical strength of headless cannulated screws about screw length using the tension band wiring technique. Forty-eight sawbone patellas with transverse fractures were divided into three fixation groups based on the screw length used in tension band wiring. Overall, three different fixation groups were determined: Group 1 (recessed headless cannulated screw fixation), Group 2 (full-length headless cannulated screw fixation), and Group 3 (protruding headless cannulated screw fixation). A setup was used to simulate a knee with a flexion angle of 60 degrees. Specimens underwent biomechanical testing under axial traction (static test) and cyclic loading (dynamic test). Displacements at 300 Newtons (N), loads at 2 millimetres (mm) displacement, and failure loads were documented for each sample in the static test. In the dynamic test protocol, 10,000 repetitive cycles were performed under physiological load between 100 and 300 N, and final displacements were recorded. There were significant differences in the loads achieved at 2 mm displacement levels, and Group 3 demonstrated lower force values compared to other constructs in the static test (P = 0.003). All groups revealed similar displacements at 300 N and failure load values under axial traction. In the dynamic test, Group 3 had significantly higher fracture displacement under cyclic loading compared to the other specimens (P = 0.006). This study found headless cannulated screws for transverse patella fracture fixation provide sufficient stability; however, protruding headless screws reduce the fixation strength. Recessed or full-length screws may improve stability and bony healing, potentially preventing complications in patella fractures. Biomechanical study N/A.Keywords
This publication has 41 references indexed in Scilit:
- Comminuted patellar fractures: The role of biplanar fixed angle plate constructsJournal of Orthopaedic Translation, 2021
- The influence of gap size on the development of fracture union with a micro external fixatorJournal of the Mechanical Behavior of Biomedical Materials, 2019
- Biomechanical evaluation of the tension band wiring principle. A comparison between two different techniques for transverse patella fracture fixationInjury, 2017
- Non-metallic implant for patellar fracture fixation: A systematic reviewInjury, 2016
- Biomechanical cadaver testing of a fixed-angle plate in comparison to tension wiring and screw fixation in transverse patella fracturesInjury, 2012
- Cyclic long‐term loading of a bilateral fixed‐angle plate in comparison with tension band wiring with K‐wires or cannulated screws in transverse patella fracturesKnee Surgery, Sports Traumatology, Arthroscopy, 2012
- Double fixation of displaced patella fractures using bioabsorbable cannulated lag screws and braided polyester suture tension bandsInjury, 2011
- Cannulated Screw and Cable are Superior to Modified Tension Band in the Treatment of Transverse Patella FracturesClinical Orthopaedics and Related Research, 2011
- Fixed-angle plate osteosynthesis of the patella – An alternative to tension wiring?Clinical Biomechanics, 2010
- Anatomical and biomechanical evaluation of the tension band technique in patellar fracturesInternational Orthopaedics, 2008