Verification of a novel method for tube voltage constancy measurement of orthovoltage x‐ray irradiators
- 16 July 2014
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 41 (8Part1) , 084101
Abstract
Purpose: For orthovoltage x‐ray irradiators, the tube voltage is one of the most fundamental system parameters as this directly relates to the dosimetry in radiation biology studies; however, to the best of our knowledge, there is no commercial portable quality assurance (QA) tool to directly test the constancy of the tube voltage greater than 160 kV. The purpose of this study is to establish the Beam Quality Index (BQI), a quantity strongly correlated to the tube voltage, as an alternative parameter for the verification of the tube voltage as part of the QA program of orthovoltage x‐ray irradiators.Methods: A multipurpose QA meter and its associated data acquisition software were used to customize the measurement parameters to measure the BQI and collect its time‐plot. BQI measurements were performed at 320 kV with four filtration levels on three orthovoltage x‐ray irradiators of the same model, one of which had been recently energy‐calibrated at the factory.Results: For each of the four filtration levels, the measured BQI values were in good agreement (<5%) between the three irradiators. BQI showed filtration‐specificity, possibly due to the difference in beam quality.Conclusions: The BQI has been verified as a feasible alternative for monitoring the constancy of the tube voltage for orthovoltage irradiators. The time‐plot of BQI offers information on the behavior of beam energy at different phases of the irradiation time line. In addition, this would provide power supply performance characteristics from initial ramp‐up to plateau, and finally, the sharp drop‐off at the end of the exposure.Keywords
Funding Information
- National Institute of Allergy and Infectious Diseases (NIAID) (5U19AI067798)
- U.S. Nuclear Regulatory Commission (NRC) (NRC-HQ-12-G-38-0022)
- U.S. Nuclear Regulatory Commission (NRC‐HQ‐12‐G‐38‐0022)
- National Institute of Allergy and Infectious Diseases (5U19AI067798)
- U.S. Nuclear Regulatory Commission (NRC-HQ-12-G-38-0022)
- National Institute of Allergy and Infectious Diseases (5U19AI067798)
This publication has 5 references indexed in Scilit:
- Specific issues in small animal dosimetry and irradiator calibrationInternational Journal of Radiation Biology, 2011
- Geant4 toolkit for simulation of HEP experimentsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Simplified method for standardization of roentgenographic image quality using penetration quality measurementsAmerican Journal of Roentgenology, 1976
- Measurement of kilovoltage by the penetrameter-theoretical aspects (X-ray generators)Physics in Medicine & Biology, 1975
- Checking diagnostic X-ray beam qualityBritish Journal of Radiology, 1968