Abstract
The pullout strengths of various pedicle screw designs are compared using tapped and untapped pilot holes. The objective of this study is to compare the pullout strength of various pedicle screw designs. The designs are compared using tapped and untapped pilot holes. By using several different screw designs, it is possible to gain an understanding of whether there is a correlation between tapping a pilot hole and the ultimate pullout strength. Most bone screws originally developed were intended to be installed in a pretapped pilot hole. This same technology has been carried over to the development of more modern bone screws for use in spinal fixation applications. Many pedicle screws in use today are still intended to be installed in a tapped hole. Preparing the vertebrae and tapping of a pilot hole involve additional trauma to the patient as well as increased operating time. Pedicle screws from various manufacturers are installed in tapped and untapped pilot holes and then loaded to failure. A uniform synthetic material was used to provide a consistent test of each screw design by eliminating variability seen in bone. Tapping pilot holes did not increase the pullout strength of the screws tested in this study. It was observed during testing that tapping some of the holes degraded the material. This degradation led to pullout strengths that were lower than in the untapped case, and generally larger standard deviations. The pullout strength was not increased by tapping for the screws in this study. Screws placed in untapped holes generally had higher pullout strengths and lower standard deviations. The results of this study suggest that tapping does not increase pullout strength in bone with densities near 20 lb/ft3, which correlates with low density cancellous or osteoporotic bone.Keywords
This publication has 13 references indexed in Scilit:
- Increase of pullout strength of spinal pedicle screws with conical core: Biomechanical tests and finite element analysesJournal of Orthopaedic Research, 2005
- Biomechanical Evaluation of Relationship of Screw Pullout Strength, Insertional Torque, and Bone Mineral Density in the Cervical SpineJournal of Spinal Disorders & Techniques, 2004
- Revision of Failed Pedicle Screws Using Hydroxyapatite CementSpine, 1998
- Comparative Pull-Out Strength of Tapped and Untapped Pilot Holes for Bicortical Anterior Cervical ScrewsSpine, 1997
- Factors Affecting the Pullout Strength of Cancellous Bone ScrewsJournal of Biomechanical Engineering, 1996
- Screw Fixation in the Human SacrumSpine, 1992
- Implant holding power of the femoral headActa Orthopaedica, 1992
- Optimizing Bone Screw Pullout ForceJournal of Orthopaedic Trauma, 1990
- Torsion and bending analysis of internal fixation techniques for femoral neck fractures: The role of implant design and bone densityJournal of Orthopaedic Research, 1987
- The mechanical properties of surgical bone screws and some aspects of insertion practiceInjury, 1972