Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal‐on‐metal total hip arthroplasty system
- 25 July 2015
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 33 (12) , 1868-1874
Abstract
Taper wear at the head‐neck junction is a possible cause of early failure in large head metal‐on‐metal (LH‐MoM) hip replacements. We hypothesized that: (i) taper wear may be more pronounced in certain product designs; and (ii) an increased abductor moment arm may be protective. The tapers of 104 explanted LH‐MoM hip replacements revised for adverse reaction to metal debris (ARMD) from a single manufacturer were analyzed for linear and volumetric wear using a co‐ordinate measuring machine. The mated stem was a shorter 12/14, threaded trunnion (n = 72) or a longer, smooth 11/13 trunnion (n = 32). The abductor moment arm was calculated from pre‐revision radiographs. Independent predictors of linear and volumetric wear included taper angle, stem type, and the horizontal moment arm. Tapers mated with the threaded 12/14 trunnion had significantly higher rates of volumetric wear (0.402 mm3/yr vs. 0.123 mm3/yr [t = −2.145, p = 0.035]). There was a trend to larger abductor moment arms being protective (p = 0.055). Design variation appears to play an important role in taper‐trunnion junction failure. We recommend that surgeons bear these findings in mind when considering the use of a short, threaded trunnion with a cobalt‐chromium head. © 2015 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 33:1868–1874, 2015.Keywords
Funding Information
- Food and Drug Administration (HHSF223201300925P)
This publication has 15 references indexed in Scilit:
- Why does titanium alloy wear cobalt chrome alloy despite lower bulk hardness: A nanoindentation study?Thin Solid Films, 2013
- Enhanced wear and corrosion in modular tapers in total hip replacement is associated with the contact area and surface topographyJournal of Orthopaedic Research, 2013
- Material loss at the taper junction of retrieved large head metal‐on‐metal total hip replacementsJournal of Orthopaedic Research, 2013
- Trunnionosis: A pain in the neckWorld Journal of Orthopedics, 2013
- Taper junction failure in large-diameter metal-on-metal bearingsBone & Joint Research, 2012
- Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and WalesThe Lancet, 2012
- Volumetric wear assessment of failed metal-on-metal hip resurfacing prosthesesWear, 2011
- Accelerating failure rate of the ASR total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2011
- Large Diameter Femoral Heads Impose Significant Alterations on the Strains Developed on Femoral Component and Bone: A Finite Element AnalysisThe Open Orthopaedics Journal, 2011
- High failure rates with a large-diameter hybrid metal-on-metal total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2011