A Finite Element Analysis of a New Dental Implant Design: The Influence of the Diameter, Length, and Material of an Implant on Its Biomechanical Behavior
Open Access
Abstract
It is widely recognized that excessive stress and/or strain can lead to peri-implant bone atrophy; therefore, the clinical success of dental implants is intrinsically related to their biomechanical behavior. This study evaluates the influence of the diameter, length, and material [Ti6Al4V (α+β Ti) and Ti35Nb7Zr5Ta (β-Ti)] of a novel cylindrical dental implant on stress and strain levels within maxillary bone of type II quality. The implant design aims to ensure an appropriate distribution of stresses and strains within the peri-implant bone structures (cortical and trabecular bones) while also facilitating surgical machining by requiring a simple, linear, and less expensive bone incision. This approach minimizes the risk of thermal necrosis, a common complication in osteotomies for conical implants that can lead to peri-implant bone loss. Using finite element analysis, stress and strain patterns were evaluated in the maxillary second premolar region under static delayed loading. The results reveal that the cortical bone strains remained below the critical threshold (0.003) to prevent resorption. In the trabecular bone, only larger diameter/length configurations satisfied the previous strain criterion. In all simulations, trabecular bone stress remained below 3 MPa, whereas cortical bone stress peaked at 78 MPa. Notably, the implant model with the largest diameter/length minimized stress and strain concentrations in type II bone when compared to smaller designs, thereby demonstrating its biomechanical advantage.Keywords
Funding Information
- MICIU/AEI/10.13039/501100011033 (PDC2022-133369-I00, PN385LH007-035: 9827)
- ERDF/EU (PDC2022-133369-I00, PN385LH007-035: 9827)
- Ministerio de Ciencia e Innovación (PDC2022-133369-I00, PN385LH007-035: 9827)
- 9827 of the Cuban National Science and Technology program: Bio-technology, Pharmaceutical Industry, and Medical Technologies (PDC2022-133369-I00, PN385LH007-035: 9827)
This publication has 72 references indexed in Scilit:
- Finite element analysis in implant dentistry: State of the art and future directionsDental Materials, 2023
- Relationship between dental implant macro-design and osseointegration: a systematic reviewOral and Maxillofacial Surgery, 2022
- Short versus standard implants at sinus augmented sites: a systematic review and meta-analysisClinical Oral Investigations, 2022
- Effects of assessing the bone remodeling process in biomechanical finite element stability evaluations of dental implantsComputer Methods and Programs in Biomedicine, 2022
- Biomechanical effects of dental implant diameter, connection type, and bone density on microgap formation and fatigue failure: A finite element analysisComputer Methods and Programs in Biomedicine, 2021
- Effect of bone quality and quantity on the primary stability of dental implants in a simulated bicortical placementClinical Oral Investigations, 2020
- Effect of implant macrogeometry on peri-implant healing outcomes: a randomized clinical trialClinical Oral Investigations, 2018
- Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element AnalysisThe Journal of Indian Prosthodontic Society, 2012
- A mathematical model for simulating the bone remodeling process under mechanical stimulusDental Materials, 2007
- Bone's mechanostat: A 2003 updateThe Anatomical Record Part A: Discoveries in Molecular, Cellular, and Evolutionary Biology, 2003