Real-Time Monitoring System for Thermocouple Inhomogeneity Equipment
Authors:
Open Access
- 1 June 2022
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 2243 (1) , 012039
Abstract
Thermocouple inhomogeneity equipment was developed to measure the Seebeck coefficient characteristic in each wire position based on the effect of temperature exposure along the thermocouple wire. Temperature and distance sensors are used to measure the temperature output of the heating system and the movement of the thermocouple position. We developed a real-time measurement system to monitor the measurement of the thermocouple sensor based on a microcontroller. This monitoring system is needed to monitor changes in the ice bath during measurement, if the ice bath was unstable or being melting, it can affect the output voltage of the thermocouple. The program was compiled and uploaded to ESP 32 then monitoring data with an open-source internet of things called Thingspeak was used as a platform to retrieve and display the collected data. Real-time monitoring of distance from ultrasonic and temperature of heat source (heat gun) can be accessed through web applications and smartphones. The data were analyzed based on conditions for Seebeck value of inhomogeneity and validated with standard calibration using the liquid bath. This study has been considered successfully implemented that performs on IoT platforms and was adaptively ready to monitor thermocouple inhomogeneity equipment.Keywords
This publication has 3 references indexed in Scilit:
- Smart irrigation system based on internet of things (IOT)Journal of Physics: Conference Series, 2019
- Server Room Temperature & Humidity Monitoring Based on Internet of Thing (IoT)Journal of Physics: Conference Series, 2019
- Thermal Preconditioning of MIMS Type K Thermocouples to Reduce DriftInternational Journal of Thermophysics, 2016