Effect of Surgical Technique on Electrode Impedance after Cochlear Implantation
Authors:
- Frédéric VenailFVFrédéric Venail
- Jean-Pierre PironJPJean-Pierre Piron
- Charlène BatrelCBCharlène Batrel
- Pierfrancesco PellicciaPPPierfrancesco Pelliccia
- Françoise ArtièresFAFrançoise Artières
- Alain UzielAUAlain Uziel
- Michel MondainMMMichel Mondain
- 1 August 2011
- journal article
- research article
- Published by SAGE Publications in Annals of Otology, Rhinology & Laryngology
- Vol. 120 (8) , 529-534
Abstract
Objectives: We compare the evolution of electrode impedance values (IVs) following either conventional cochlear implantation or implantation by the soft surgery (SS) technique. Methods: We performed a retrospective chart review of 20 consecutive adult patients who underwent implantation with the Nucleus CA 24 device between 2004 and 2007. Five patients with preoperative residual hearing at the frequencies 256, 512, and 1,024 Hz underwent implantation by an SS cochlear implantation technique (SS group), and the 15 other patients underwent a conventional implantation technique (conventional cochleostomy [CC] group). The active electrodes were classified as distal (17 to 22), middle (10 to 16), or proximal (3 to 9) according to their position in relation to the tip of the electrode array. Their IVs were collected at 1, 3, 12, 24, and 36 months after implantation. Changes in auditory thresholds at 3 and 24 months were reported for patients in the SS group. Results: The postoperative IVs of both the CC and SS groups decreased significantly between 1 and 3 months after implantation (p < 0.05) and then remained stable. The IVs after 12 months were significantly lower (p < 0.05) in the SS group than in the CC group. Conclusions: Patients who underwent the SS technique displayed lower long-term electrode IVs than did their counterparts in the CC group. If electrode IVs are indeed an indirect representation of cochlear fibrosis, the use of the SS technique in lieu of the CC technique could reduce fibrotic development.Keywords
This publication has 16 references indexed in Scilit:
- The Injured Cochlea as a Target for Inflammatory Processes, Initiation of Cell Death Pathways and Application of Related Otoprotective StrategiesRecent Patents on CNS Drug Discovery, 2010
- Histopathologic Assessment of Fibrosis and nEw Bone Formation in Implanted Human Temporal Bones Using 3D ReconstructionOtolaryngology–Head and Neck Surgery, 2009
- Analysis of Intracochlear New Bone and Fibrous Tissue Formation in Human Subjects with Cochlear ImplantsAnnals of Otology, Rhinology & Laryngology, 2007
- Prevention of cochlear implant electrode damageCurrent Opinion in Otolaryngology & Head and Neck Surgery, 2006
- Dual Electrode Stimulation Using the Nucleus CI24RE Cochlear Implant: Electrode Impedance and Pitch Ranking StudiesEar & Hearing, 2005
- Predicting the effect of post-implant cochlear fibrosis on residual hearingHearing Research, 2005
- A Longitudinal Study of Electrode Impedance, the Electrically Evoked Compound Action Potential, and Behavioral Measures in Nucleus 24 Cochlear Implant UsersEar & Hearing, 2001
- Chronic electrical stimulation of the auditory nerve using high surface area (HiQ) platinum electrodesHearing Research, 2001
- Intracochlear Factors Contributing to Psychophysical Percepts Following Cochlear ImplantationActa Oto-Laryngologica, 1998
- Chronic electrical stimulation of the auditory nerve at high stimulus rates: a physiological and histopathological studyHearing Research, 1997