Development of a dual‐spray electrospray deposition system for matrix‐assisted laser desorption/ionization time‐of‐flight mass spectrometry
Authors:
- 12 March 2008
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 22 (8) , 1168-1174
Abstract
A new method of matrix‐assisted laser desorption/ionization (MALDI) sample preparation using a dual‐spray electrospray deposition system is demonstrated and employed for the investigation of gas‐phase cationization reactions in the MALDI plume. The dual‐spray electrospray system is found to increase the homogeneity of the sample similarly to that of a conventional single‐spray electrospray system. The dual‐spray electrospray system allows for intimate mixing of separately prepared sample components and results in improved quantitative results. The development of this device also leads to the possibility of mixing sample components prepared in different solvents without the need to be concerned with solvent miscibility. Copyright © 2008 John Wiley & Sons, Ltd.Keywords
This publication has 24 references indexed in Scilit:
- Three-layer matrix/sample preparation method for MALDI MS analysis of low nanomolar protein samplesJournal of the American Society for Mass Spectrometry, 2006
- Updating Evidence for Cationization of Polymers in the Gas Phase during Matrix-Assisted Laser Desorption/IonizationEuropean Journal of Mass Spectrometry, 2005
- MALDI-TOF-MS Analysis of Droplets Prepared in an Electrodynamic Balance: “Wall-less” Sample PreparationAnalytical Chemistry, 2001
- Poly(ethylene glycol) limits of detection using internal matrix-assisted laser desorption/ionization fourier transform mass spectrometryJournal of Mass Spectrometry, 1998
- Influence of Matrix Solution Conditions on the MALDI-MS Analysis of Peptides and ProteinsAnalytical Chemistry, 1996
- Low-flow interface for liquid chromatography–inductively coupled plasma mass spectrometry speciation using an oscillating capillary nebulizerJournal of Analytical Atomic Spectrometry, 1996
- Analysis of Single Mammalian Cell Lysates by Mass SpectrometryJournal of the American Chemical Society, 1996
- Improved Resolution and Very High Sensitivity in MALDI TOF of Matrix Surfaces Made by Fast EvaporationAnalytical Chemistry, 1994
- A method to increase contaminant tolerance in protein matrix‐assisted laser desorption/ionization by the fabrication of thin protein‐doped polycrystalline filmsRapid Communications in Mass Spectrometry, 1994
- Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltonsAnalytical Chemistry, 1988