Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection
Open Access
- 26 May 2012
- journal article
- research article
- Published by Springer Nature in The International Journal of Cardiovascular Imaging
- Vol. 29 (1) , 39-51
Abstract
Recently, Fourier domain OCT (FD-OCT) has been introduced for clinical use. This approach allows in vivo, high resolution (15 micron) imaging with very fast data acquisition, however, it requires brief flushing of the lumen during imaging. The reproducibility of such fast data acquisition under intracoronary flush application is poorly understood. To assess the inter-study variability of FD-OCT and to compare lumen morphometry to the established invasive imaging method, IVUS. 18 consecutive patients with coronary artery disease scheduled for PCI were included. In each target vessel a FD-OCT pullback (MGH system, light source 1,310 nm, 105 fps, pullback speed 20 mm/s) was acquired during brief (3 s) injection of X-ray contrast (flow 3 ml/s) through the guiding catheter. A second pullback was repeated under the same conditions after re-introduction of the FD OCT catheter into the coronary artery. IVUS and OCT imaging was performed in random order. FD-OCT and IVUS pullback data were analyzed using a recently developed software employing semi automated lumen contour and stent strut detection algorithms. Corresponding ROI were matched based on anatomical landmarks such as side branches and/or stent edges. Inter-study variability is presented as the absolute difference between the two pullbacks. FD-OCT showed remarkably good reproducibility. Inter-study variability in native vessels (cohort A) was very low for mean and minimal luminal area (0.10 ± 0.38, 0.19 ± 0.57 mm2, respectively). Likewise inter-study variability was very low in stented coronary segments (cohort B) for mean lumen, mean stent, minimal luminal and minimal stent area (0.06 ± 0.08, 0.07 ± 0.10, 0.04 ± 0.09, 0.04 ± 0.10 mm2, respectively). Comparison to IVUS morphometry revealed no significant differences. The differences between both imaging methods, OCT and IVUS, were very low for mean lumen, mean stent, minimal luminal and minimal stent area (0.10 ± 0.45, 0.10 ± 0.36, 0.26 ± 0.54, 0.05 ± 0.47 mm2, respectively). FD-OCT shows excellent reproducibility and very low inter-study variability in both, native and stented coronary segments. No significant differences in quantitative lumen morphometry were observed between FD-OCT and IVUS. Evaluating these results suggest that FD-OCT is a reliable imaging tool to apply in longitudinal coronary artery disease studies.Keywords
This publication has 36 references indexed in Scilit:
- In vitro validation of new Fourier-domain optical coherence tomographyEuroIntervention, 2011
- Second-Generation Optical Coherence Tomography in Clinical Practice. High-Speed Data Acquisition is Highly Reproducible in Patients Undergoing Percutaneous Coronary InterventionRevista Española de Cardiología (English Edition), 2010
- Reproducibility of coronary Fourier domain optical coherence tomography: quantitative analysis of in vivo stented coronary arteries using three different software packagesEuroIntervention, 2010
- An optical coherence tomography study of a biodegradable vs. durable polymer-coated limus-eluting stent: a LEADERS trial sub-studyEuropean Heart Journal, 2009
- Three-Dimensional Coronary Artery Microscopy by Intracoronary Optical Frequency Domain ImagingJACC: Cardiovascular Imaging, 2008
- In Vivo Comparison Between Optical Coherence Tomography and Intravascular Ultrasound for Detecting Small Degrees of In-Stent Neointima After Stent ImplantationJACC: Cardiovascular Interventions, 2008
- Comprehensive volumetric optical microscopy in vivoNature Medicine, 2006
- In vivo Variability in Quantitative Coronary Ultrasound and Tissue Characterization Measurements with Mechanical and Phased-array CathetersThe International Journal of Cardiovascular Imaging, 2005
- Intracoronary ultrasound imaging: Intraobserver and interobserver variability of morphometric measurementsAmerican Heart Journal, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986