The revision rate of large head metal-on-metal and resurfacing hips are significantly higher than conventional total hip replacements. The revision of these components has been linked to high wear caused by edge loading; which occurs when the head–cup contact patch extends over the cup rim. There are two current explanations for this; first, there is loss of entrainment of synovial fluid resulting in breakdown of the lubricating film and second, edge loading results in a large local increase in contact pressure and consequent film thickness reduction at the cup rim, which causes an increase in wear. This paper develops a method to calculate the distance between the joint reaction force vector and the cup rim – the contact patch centre to rim (CPCR) distance. However, the critical distance for the risk of edge loading is the distance from the contact patch edge to rim (CPER) distance. An analysis of explanted hip components, divided into edge worn and non-edge-worn components showed that there was no statistical difference in CPCR values, but the CPER value was significantly lower for edge worn hips. Low clearance hips, which have a more conformal contact, have a larger diameter contact patch and thus are more at risk of edge loading for similarly positioned hips.Keywords
This publication has 30 references indexed in Scilit:
- Bioengineering reasons for the failure of metal-on-metal hip prosthesesThe Journal of Bone and Joint Surgery. British volume, 2011
- Lubricating film thickness measurements with bovine serumTribology International, 2011
- Adverse reaction to metal debris following hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2011
- Hard-on-Hard Total Hip Impingement Causes Extreme Contact Stress ConcentrationsClinical Orthopaedics and Related Research, 2011
- Analysis of wear of retrieved metal-on-metal hip resurfacing implants revised due to pseudotumoursThe Journal of Bone and Joint Surgery. British volume, 2010
- Wear mechanisms in metal-on-metal bearings: The importance of tribochemical reaction layersJournal of Orthopaedic Research, 2009
- High Cup Angle and Microseparation Increase the Wear of Hip Surface ReplacementsClinical Orthopaedics and Related Research, 2009
- Ten different hip resurfacing systems: biomechanical analysis of design and material propertiesInternational Orthopaedics, 2008
- The John Charnley AwardClinical Orthopaedics and Related Research, 2006
- The Tribology (Friction, Lubrication and Wear) of All-Metal Artificial Hip JointsClinical Orthopaedics and Related Research, 1996