Technical note: Low clinical efficacy, but good acceptability of a point‐of‐care electronic palpation device for breast cancer screening for a lower middle‐income environment
Authors:
- 10 February 2022
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 49 (4) , 2663-2671
Abstract
Background : Late-stage breast cancer rates in the Pacific where mammography services are limited are exceedingly high: Marshall Islands (61%), Palau (94%), and Samoa (79%). Due to the limited medical resources in these areas an alternative accessible technology is needed. The iBreast Exam (iBE) is a point-of-care electronic palpitation device that has a reported sensitivity of 86%. However, little is known about the performance and acceptability of this device for women in the Pacific. Methods : A total of 39 women (ages 42–73 years) were recruited in Guam with 19 women having a mammogram requiring biopsy (BI-RADS category 4 or above) and 20 women with a negative screening mammogram before study visit. Participants received an iBE exam and completed a 26-item breast health questionnaire to evaluate the iBE. Furthermore, the performance characteristics of the iBE were tested using gelatin breast phantoms in terms of tumor size, tumor depth, and overall breast stiffness. Results : The iBE had a sensitivity of 20% (2 true positives to 8 false negatives) and specificity of 92% (24 false positives to 278 true negatives) when analyzed based on location of tumor by quadrant. The iBE also had generally poor agreement according to a Cohen's kappa value of 0.068. The phantom experiments showed that the iBE can detect tumors as deep as 2.5cm, but only if the lesion is greater than 8mm in diameter. However, the iBE did demonstrate acceptability; 67% of the women reported that they had high trust in iBE as an early detection device. Conclusions : The iBE had generally poor sensitivity and specificity when tested in a clinical setting which does not allow its use as a screening tool. Impact : This study demonstrates the need for an alternative screening method other than electronic palpation for lower-middle income areas. This article is protected by copyright. All rights reservedKeywords
Funding Information
- National Institutes of Health (NIH U54, U54CA143727)
This publication has 17 references indexed in Scilit:
- Technical Note: Multipurpose CT, ultrasound, and MRI breast phantom for use in radiotherapy and minimally invasive interventionsMedical Physics, 2016
- Development of Oil-in-Gelatin Phantoms for Viscoelasticity Measurement in Ultrasound Shear Wave ElastographyUltrasound in Medicine & Biology, 2014
- Breast Tumor Detection Using Piezoelectric Fingers: First Clinical ReportJournal of the American College of Surgeons, 2013
- Epidemiology, screening and diagnosis of breast cancer in the Asia–Pacific region: Current perspectives and important considerationsAsia-Pacific Journal of Clinical Oncology, 2008
- Breast Cancer Screening Policies in Developing Countries: A Cost-effectiveness Analysis for IndiaJNCI: Journal of the National Cancer Institute, 2008
- Developing a Questionnaire to Measure Perceived Attributes of eHealth InnovationsAmerican Journal of Health Behavior, 2007
- Elastic moduli of normal and pathological human breast tissues: an inversion-technique-based investigation of 169 samplesPhysics in Medicine & Biology, 2007
- An inverse problem solution for measuring the elastic modulus of intactex vivobreast tissue tumoursPhysics in Medicine & Biology, 2007
- Revised susceptibility, benefits, and barriers scale for mammography screeningResearch in Nursing & Health, 1999
- A Coefficient of Agreement for Nominal ScalesEducational and Psychological Measurement, 1960