Unexpected Collection of Pulmonary Venous Blood from a Pulmonary Artery Catheter: A Case Report
Authors:
Open Access
- 23 October 2016
- journal article
- Published by Brieflands in Anesthesiology and Pain Medicine
- Vol. 6 (6) , e42621
Abstract
Background: Pressure and waveform at the catheter tip are continuously monitored during catheterization of pulmonary artery to ensure accurate catheter placement. We present a case in which pulmonary venous blood was unexpectedly collected from the pulmonary artery catheter despite pulmonary artery pressure and waveform detection at the catheter tip, and describe the measures taken to correct the catheter placement. Case Presentation: A 74-year-old male underwent mitral valve plasty for cardiac failure caused by mitral valve regurgitation. Preoperative transthoracic echocardiography showed no septal shunt. The pulmonary artery was catheterized through a sheath introducer in the right jugular vein, and the balloon was inflated after insertion of a 15-cm catheter. The catheter was advanced until a pulmonary artery waveform was detected and the pulmonary artery wedge pressure was 21 mmHg at end-expiration. The balloon was deflated and the catheter tip was pulled back 3 cm. Pulmonary artery waveforms and appropriate a and v waves were detected, and transesophageal echocardiography confirmed the location of the catheter tip in the right pulmonary artery. The first collected blood sample had an oxygen partial pressure of 358.8 mmHg, carbon dioxide partial pressure of 20.1 mmHg, and oxygen saturation of 99%, indicating pulmonary venous blood. The pulmonary artery catheter was pulled back 5 cm, but a second blood sample showed the same results. The catheter was pulled back a further 6 cm while the location of the catheter tip was monitored on X-ray fluoroscopy. Blood gas testing through the catheter tip showed oxygen saturation of 84.4 % and oxygen partial pressure of 41.6 mmHg. Surgery was performed uneventfully. Postoperative chest radiographs showed proper placement of the pulmonary artery catheter, but radiographs on postoperative day 1 showed over-insertion, although the insertion length was unchanged. The catheter was removed. The patient was discharged 2 months postoperatively. Conclusions: Our case highlights the fact that the tip of the pulmonary artery catheter can easily advance into a peripheral branch of the pulmonary artery and cause pulmonary venous blood to be sampled instead of pulmonary arterial blood. A variety of monitoring techniques are needed to confirm accurate catheter placement. Keywords: Pulmonary Artery Catheter; Blood Sampling; Pulmonary Artery; Blood Sampling; Pulmonary Artery; Case ReportKeywords
This publication has 11 references indexed in Scilit:
- Pulmonary artery pseudoaneurysm after Swan-Ganz catheterization: a case presentation and review of literatureEuropean Heart Journal: Acute Cardiovascular Care, 2014
- Saudi Guidelines on the Diagnosis and Treatment of Pulmonary Hypertension: Perioperative management in patients with pulmonary hypertensionAnnals of Thoracic Medicine, 2014
- The current state of fellowship training in pulmonary artery catheter placement and data interpretation: A national survey of pulmonary and critical care fellowship program directorsJournal of Critical Care, 2013
- Accuracy and precision of echocardiography versus right heart catheterization for the assessment of pulmonary hypertensionInternational Journal of Cardiology, 2013
- Pulmonary artery catheter coiled in the main pulmonary artery trunkBMJ Case Reports, 2013
- Pulmonary artery catheter induced pulmonary haemorrhage2009
- [A case of pulmonary artery rupture due to pulmonary artery catheter malposition after the weaning of cardiopulmonary bypass].2003
- Pulmonary Artery Rupture Associated With the Swan-Ganz CatheterChest, 1995
- A Prospective Study of Complications of Pulmonary Artery Catheterizations in 500 Consecutive PatientsChest, 1983
- Complications of Pulmonary Artery Catheterization in the Care of Critically III PatientsChest, 1979