Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations
- 15 April 2008
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 18 (8) , 1674-1682
Abstract
We evaluated the potential for reduction of dose to the female breast in computed tomography (CT) of the thorax by using three different techniques: bismuth shielding, partial CT scanning and tube-current modulation (TCM). Measurements and simulations of dose and image quality were performed for a 64-slice CT system using a semi-anthropomorphic thorax phantom with breasts added. Three-dimensional dose distributions were calculated by Monte Carlo (MC) methods. Noise was determined by measurements and simulations. Bismuth shielding resulted in a dose reduction of about 50% for the breast, noise increased up to 40% and image quality was impaired by artifacts. In partial CT scans, not irradiating the breasts directly, dose to the breasts was reduced typically by 50%. To sustain a constant noise level, an increase of irradiation in the anteroposterior position resulted in a higher dose to the spine. Reduction of dose to the breasts of about 10% was achieved with TCM; distribution of noise was homogeneous and image quality uniform. Reduction of dose to the female breast was achieved by using all adapted CT methods. Bismuth shielding may compromise image quality, increase noise level and introduce streak artifacts. Partial and TCM examinations reduced dose to the breast without influencing image quality.Keywords
This publication has 15 references indexed in Scilit:
- Validation of a Monte Carlo tool for patient-specific dose simulations in multi-slice computed tomographyEuropean Radiology, 2007
- Coronary Artery Calcium: A Multi-institutional, Multimanufacturer International Standard for Quantification at Cardiac CTRadiology, 2007
- Urinary Stone Disease: Comparison of Standard-Dose and Low-Dose with 4D MDCT Tube Current ModulationAmerican Journal of Roentgenology, 2007
- Dose Reduction and Image Quality in MDCT Colonography Using Tube Current ModulationAmerican Journal of Roentgenology, 2006
- Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image qualityEuropean Radiology, 2006
- Use of an Automatic Exposure Control Mechanism for Dose Optimization in Multi–Detector Row CT Examinations: Clinical EvaluationRadiology, 2005
- Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the users' needs?Clinical Radiology, 2004
- Strategies for CT Radiation Dose OptimizationRadiology, 2004
- Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadaversEuropean Radiology, 1999
- Quantitative CT of the Lung with Spirometrically Controlled Respiratory Status and Automated Evaluation Procedures1990