Repeatability of a primary thermometer and calibration of resistance thermometers between 5 K and 25 K
- 12 December 2024
- journal article
- research article
- Published by IOP Publishing in Metrologia
- Vol. 62 (1) , 015005
Abstract
This work describes improvements to our Single Pressure Refractive-Index Gas Thermometry (SPRIGT) system (Gao et al 2020 Metrologia 57 065006), where a new experimental procedure for Constant Refractive-Index Gas Thermometry (CRIGT) was proposed and implemented, and after 2.5 years new values of T–T 90 obtained in the temperature range from 5 K to 25 K. Comparison results indicate that our SPRIGT system is robust with excellent repeatability, reliability, and stability. Not only is the uncertainty in thermodynamic temperature T still below 0.17 mK, but also good reproducibility of T 90 (temperature of the 1990 International Temperature Scale), T and T–T 90 is observed with values of 21 μK, 28 μK and 35 μK, respectively. Furthermore, we have implemented direct dissemination of thermodynamic temperature T by establishing our own thermodynamic temperature wire scale under the new International System of Units (SI) from 5 K to 25 K. The remarkable agreement further enhances our confidence on SPRIGT for implementing the redefined kelvin below 25 K, disseminating thermodynamic temperature T, and ultimately partaking in an international comparison of thermodynamic temperature via highly stable resistance thermometers.Keywords
Funding Information
- National Key Research and Development Program of China (2022YFE0210200)
- National Natural Science Foundation of China (52125602 and 52006231)
- International Partnership Program of the Chinese Academy of Sciences (1A1111KYSB20210024)
- Scientific Instrument Developing Project of the Chinese Academy of Sciences (ZDKYYQ20210001)
- Youth Innovation Promotion Association of the Chinese Academy of Sciences (2022028)
- Beijing Municipal Science & Technology Commission, Administrative Commission of Zhongguancun Science Park (Z221100002722005)
- CAS President’s International Fellowship Initiative (2019VEA0015)
- Direk-T (22IEM02)
This publication has 29 references indexed in Scilit:
- 2022 Update for the Differences Between Thermodynamic Temperature and ITS-90 Below 335 KJournal of Physical and Chemical Reference Data, 2022
- Investigation of High-Stability Temperature Control in Primary Gas ThermometryJournal of Thermal Science, 2022
- Realization of ppm level pressure stability for primary thermometry using a primary piston gaugeMeasurement, 2020
- Thermal response characteristics of a SPRIGT primary thermometry systemCryogenics, 2019
- Realization of an ultra-high precision temperature control in a cryogen-free cryostatReview of Scientific Instruments, 2018
- Characteristics of Standard Capsule-Type PtCo Resistance Thermometers Between 0.65 K and 25 KInternational Journal of Thermophysics, 2017
- Method for extrapolating the compressibility data of solids from room to lower temperaturesPhysica Status Solidi (b), 2016
- Path-integral Mayer-sampling calculations of the quantum Boltzmann contribution to virial coefficients of helium-4The Journal of Chemical Physics, 2012
- Path-integral calculation of the third virial coefficient of quantum gases at low temperaturesThe Journal of Chemical Physics, 2011
- Final Report on CCT-K1: Realizations of the ITS-90, 0.65 K to 24.5561 K, using rhodium–iron resistance thermometersMetrologia, 2006