A randomized controlled trial comparing tibial migration of the ATTUNE cemented cruciate-retaining knee prosthesis with the PFC-sigma design
Authors:
Open Access
- 1 September 2020
- journal article
- research article
- Published by British Editorial Society of Bone & Joint Surgery in The Bone & Joint Journal
- Vol. 102-B (9) , 1158-1166
Abstract
Aims The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). Methods A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively. Results The overall migration (mean maximum total point motion (MPTM)) at two years was comparable: mean 1.13 mm (95% confidence interval (CI), 0.97 to 1.30) for the ATTUNE and 1.16 mm (95% CI, 0.99 to 1.35) for the PFC-sigma. At two years, the mean backward tilting was -0.43° (95% CI, -0.65 to -0.21) for the ATTUNE and 0.08° (95% CI -0.16 to 0.31), for the PFC-sigma. Overall migration between the first and second postoperative year was negligible for both components. The clinical outcomes and PROMs improved compared with preoperative scores and were not different between groups. Radiolucencies at the implant-cement interface were mainly seen below the medial baseplate: 17% in the ATTUNE and 3% in the PFC-sigma at two weeks, and at two years 42% and 9% respectively (p = 0.001). Conclusion In the first two postoperative years the initial version of the ATTUNE tibial component was not inferior with respect to overall migration, although it showed relatively more backwards tilting and radiolucent lines at the implant-cement interface than the PFC-sigma. The version of the ATTUNE tibial component examined in this study has subsequently undergone modification by the manufacturer. Level of Evidence: 1 (randomized controlled clinical trial) Cite this article: Bone Joint J 2020;102-B(9):1158–1166.Keywords
This publication has 39 references indexed in Scilit:
- Early migration of tibial components is associated with late revisionActa Orthopaedica, 2012
- RSA prediction of high failure rate for the uncoated Interax TKA confirmed by meta-analysisActa Orthopaedica, 2012
- Alternative analyses for handling incomplete follow-up in the intention-to-treat analysis: the randomized controlled trial of balloon kyphoplasty versus non-surgical care for vertebral compression fracture (FREE)BMC Medical Research Methodology, 2012
- A Clinical and Radiostereometric Study of the Cemented PFC-Sigma Prosthesis –A 5-Year Study of 29 Cases with a Fixed BearingThe Journal of Knee Surgery, 2009
- Artifacts Found During Quality Assurance Testing of Computed Radiography and Digital Radiography DetectorsJournal of Digital Imaging, 2008
- The use of the Oxford hip and knee scoresThe Journal of Bone and Joint Surgery. British volume, 2007
- Projections of Primary and Revision Hip and Knee Arthroplasty in the United States from 2005 to 2030Journal of Bone and Joint Surgery, 2007
- Guidelines for standardization of radiostereometry (RSA) of implantsActa Orthopaedica, 2005
- Quality assurance of joint replacement Legal regulation and medical judgementActa Orthopaedica, 1992
- Coxarthrosis on the island of GotlandActa Orthopaedica, 1992