Comparison of Pentacam Scheimpflug Camera with Ultrasound Pachymetry and Noncontact Specular Microscopy in Measuring Central Corneal Thickness
Authors:
- Makoto NakamuraMNMakoto Nakamura
- Yasuko TatsumiYTYasuko Tatsumi
- Azusa KusuharaAKAzusa Kusuhara
- Hidetaka MaedaHMHidetaka Maeda
- Akira NegiANAkira Negi
- 1 January 2007
- journal article
- research article
- Published by Taylor & Francis in Current Eye Research
- Vol. 32 (2) , 89-94
Abstract
Although central corneal thickness (CCT) can be measured by several methods, interchangeability of different modalities has not been fully investigated. CCT is known to correlate with intraocular pressure (IOP). The aim of this study was to evaluate the agreement of Pentacam Scheimpflug system with noncontact specular microscopy (NCSM) and ultrasound (US) pachymetry in measuring CCT and the relation between IOP taken with Goldmann applanation tonometer (GAT) and the CCT measured with these three methods. The right eyes of 135 enrolled persons without antiglaucoma drug use (100 females and 35 males), who comprised 32 patients with primary open-angle glaucoma, 14 with ocular hypertension, 45 with primary angle-closure glaucoma, and 44 controls, were studied. Intermethod comparison of CCT was made by the 95% limits of agreement analysis according to Bland and Altman. Linear regression analysis was used to assess the relationship between IOP and CCT taken with each modality. The mean CCT (+/-SD) taken with Scheimpflug, US, and NCSM was 559.49 +/- 38.44 microm, 553.01 +/- 39.33 microm, and 552.04 +/- 42.95 microm, respectively. The average values of CCT taken with the three instruments were not significantly different (one-factor ANOVA; p = 0.26), although the marginal mean difference between Scheimpflug and US or NCSM was statistically significant (paired t test; p = 0.0009 and 0.005, respectively). The 95% limits of agreement were 6.47 +/- 43.21 microm between Scheimpflug and US, 7.45 +/- 58.86 microm between Scheimpflug and NCSM, and 0.98 +/- 51.69 microm between US and NCSM. There was a positive association between IOP and CCT measured with US or NCSM, whereas there was no correlation between IOP and CCT measured with Scheimpflug. Although CCT values measured with Scheimpflug, US, and NCSM are closely similar, clinicians should keep in mind that these methods are not simply interchangeable.Keywords
This publication has 37 references indexed in Scilit:
- Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetryJournal of Cataract & Refractive Surgery, 2006
- Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetryJournal of Cataract & Refractive Surgery, 2005
- The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometryBritish Journal of Ophthalmology, 2005
- Varying Effects of Corneal Thickness on Intraocular Pressure Measurements with Different TonometersEye, 2004
- Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetryEye, 2002
- Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readingsBritish Journal of Ophthalmology, 2001
- Diurnal variations in human corneal thickness.British Journal of Ophthalmology, 1996
- A clinical comparison of optical and ultrasonic pachometryActa Ophthalmologica, 1991
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Über ApplanationstonometrieOphthalmologica, 1957