An x-ray CT polymer gel dosimetry prototype: I. Remnant artefact removal
Authors:
- A JirasekAJA Jirasek
- J CarrickJCJ Carrick
- M HiltsMHM Hilts
- 1 May 2012
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (10) , 3137-3153
Abstract
In this study a new x-ray CT polymer gel dosimetry (PGD) filtering technique is presented for the removal of (i) remnant ring and streak artefacts, and (ii) 'structured' noise in the form of minute, intrinsic gel density fluctuations. It is shown that the noise present within x-ray CT PGD images is not purely stochastic (pixel by pixel) in nature, but rather is 'structured', and hence purely stochastic-based noise-removal filters fail in removing this significant, unwanted noise component. The remnant artefact removal (RAR) technique is based on a class of signal stripping (i.e. baseline-estimation) algorithms typically used in the estimation of unwanted non-uniform baselines underlying spectral data. Here the traditional signal removal algorithm is recast, whereby the 'signal' that is removed is the structured noise and remnant artefacts, leaving the desired polymer gel dose distribution. The algorithm is extended to 2D and input parameters are optimized for PGD images. RAR filter results are tested on (i) synthetic images with measured gel background images added, in order to accurately represent actual noise present in PGD images, and (ii) PGD images of a three-field gel irradiation. RAR results are compared to a top-performing noise filter (adaptive mean, AM), used in previous x-ray CT PGD studies. It is shown that, in all cases, the RAR filter outperforms the AM filter, particularly in cases where either (i) a low-dose gel image has been acquired or (ii) the signal-to-noise ratio of the PG image is low, as in the case when a low number of image averages are acquired within a given experiment. Guidelines for the implementation of the RAR filter are given.Keywords
This publication has 18 references indexed in Scilit:
- Cosolvent-free polymer gel dosimeters with improved dose sensitivity and resolution for x-ray CT dose responsePhysics in Medicine & Biology, 2011
- Polymer gel dosimetry using x-ray computed tomography: investigation of the effect of reconstruction techniqueJournal of Physics: Conference Series, 2010
- Normoxic polymer gel dosimetry using less toxic monomer of N-isopropyl acrylamide and X-ray computed tomography for radiation therapy applicationsReports of Practical Oncology and Radiotherapy, 2010
- Polymer gel dosimeters with enhanced sensitivity for use in x-ray CT polymer gel dosimetryPhysics in Medicine & Biology, 2010
- Challenges in Credentialing Institutions and Participants in Advanced Technology Multi-institutional Clinical TrialsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Adaptive mean filtering for noise reduction in CT polymer gel dosimetryMedical Physics, 2007
- Technical considerations for implementation of x-ray CT polymer gel dosimetryPhysics in Medicine & Biology, 2005
- Image filtering for improved dose resolution in CT polymer gel dosimetryMedical Physics, 2003
- Polymer gel dosimetry using x-ray computed tomography: a feasibility study4Physics in Medicine & Biology, 2000
- Polymerization kinetics of polyacrylamide gels I. Effect of different cross‐linkersElectrophoresis, 1981