Facilitating endoscopic submucosal dissection: double balloon endolumenal platform significantly improves dissection time compared with conventional technique (with video)
Authors:
Open Access
- 16 July 2018
- journal article
- research article
- Published by Springer Nature in Surgical Endoscopy
- Vol. 33 (1) , 315-321
Abstract
Background Flexible endoscopes ability to manipulate the intestinal environment is limited. As a result, complex endolumenal procedures are often technically demanding and result in long procedure times, impacting institutional resources. Single- and double-balloon add-on endoscopic devices have been employed throughout the GI tract to facilitate tissue control e.g., small bowel enteroscopy, with recent reports suggesting a possible colonic utility for complex procedures e.g., ESD. Our objective was to objectively analyze the efficacy of a new double-balloon device in performing ESD. Methods Ex vivo—12 simulated colonic lesions were created in porcine rectum using a standard 40 mm diameter template. Two categories were evaluated, standard cap technique ESD and double-balloon assisted ESD with retraction (ESD-R). Cases were performed sequentially. In vivo—Six, 40 mm lesion ESD-R’s were performed in a porcine model. The primary outcomes of this study were total procedure and dissection times. Results In ex vivo studies, the median total procedure time with the double-balloon platform was significantly shorter than the traditional ESD technique (29 ± 18 vs. 57 ± 21 min, p = 0.03). In the in vivo studies, lesions were successfully removed in a mean time of 48 min, with a dissection time of 20 min with no significant complications. Balloon-clip retraction and specimen retrieval capabilities were used in all double-balloon assisted cases. After 6 cases, times were significantly shorter (ex vivo 47 vs. 17 min; in vivo 57 vs. 27 min). Conclusions We have demonstrated the development of a unique technical ESD method facilitated by a new double-balloon device. Ex and in vivo investigation demonstrated superiority of ESD-R over the conventional ex vivo method. The DB device provided increased stability, improved visualization and tissue traction, which significantly reduced dissection time. Such an approach may increase safety, improve patient outcomes, and may prevent unnecessary surgeries for benign conditions.Keywords
This publication has 13 references indexed in Scilit:
- Exploring the Small Bowel: Update on Deep EnteroscopyCurrent Gastroenterology Reports, 2016
- Specimen retrieval method using a sliding overtube for large colorectal neoplasm following endoscopic submucosal dissectionEndoscopy, 2015
- Endoscopic submucosal dissection—current success and future directionsNature Reviews Gastroenterology & Hepatology, 2012
- Clinical risk factors for perforation during endoscopic submucosal dissection (ESD) for large-sized, nonpedunculated colorectal tumorsSurgical Endoscopy, 2011
- Factors predictive of perforation during endoscopic submucosal dissection for the treatment of colorectal tumorsEndoscopy, 2011
- A prospective, multicenter study of 1111 colorectal endoscopic submucosal dissections (with video)Gastrointestinal Endoscopy, 2010
- Endoscopic retrieval method using a small grip-seal plastic bag for large colorectal resection specimens after endoscopic submucosal dissectionEndoscopy, 2010
- Clinicopathological factors associated with clinical outcomes of endoscopic submucosal dissection for colorectal epithelial neoplasmsEndoscopy, 2009
- Work and Resources Needed for Endoscopic Resection of Large Sessile Colorectal PolypsClinical Gastroenterology and Hepatology, 2007
- Polyp retrieval after colonoscopic polypectomy: Use of the Roth Retrieval NetGastrointestinal Endoscopy, 2001