    Comparison of Heart Rate Monitoring Accuracy between Chest Strap and Vest during Physical Training and Implications on Training Decisions

    by Jakub Parak, Mikko Salonen, Tero Myllymäki, Ilkka Korhonen

    Heart rate (HR) and heart rate variability (HRV) based physiological metrics such as Excess Post-exercise Oxygen Consumption (EPOC), Energy Expenditure (EE), and Training Impulse (TRIMP) are widely utilized in coaching to monitor and optimize an athlete’s training load. Chest straps, and recently also dry electrodes integrated to special sports vests, are used to monitor HR during sports. Mechanical design, placement of electrodes, and ergonomics of the sensor affect the measured signal quality and artefacts. To evaluate the impact of the sensor mechanical design on the accuracy of the HR/HRV and further on to estimation of EPOC, EE, and TRIMP, we recorded HR and HRV from a chest strap and a vest with the same ECG sensor during supervised exercise protocol. A 3-lead clinical Holter ECG was used as a reference. Twenty-five healthy subjects (six females) participated. Mean absolute percentage error (MAPE) for HR was 0.76% with chest strap and 3.32% with vest. MAPE was 1.70% vs. 6.73% for EE, 0.38% vs. 8.99% for TRIMP and 3.90% vs. 54.15% for EPOC with chest strap and vest, respectively. Results suggest superior accuracy of chest strap over vest for HR and physiological metrics monitoring during sports.


  • The Accuracy of Atrial Fibrillation Detection from Wrist Photoplethysmography. A Study on Post-Operative Patients

    by Adrian Tarniceriu, Jarkko Harju, Zeinab Rezaei Yousefi, Antti Vehkaoja, Jakub Parak, Arvi Yli-Hankala, Ilkka Korhonen

    Atrial fibrillation (AF) is the most common type of cardiac arrhythmia. Although not life-threatening itself, AF significantly increases the risk of stroke and myocardial infarction. Current tools available for screening and monitoring of AF are inadequate and an unobtrusive alternative, suitable for long-term use, is needed. This paper evaluates an atrial fibrillation detection algorithm based on wrist photoplethysmographic (PPG) signals. 29 patients recovering from surgery in the post-anesthesia care unit were monitored. 15 patients had sinus rhythm (SR, 67.5± 10.7 years old, 7 female) and 14 patients had AF (74.8± 8.3 years old, 8 female) during the recordings. Inter-beat intervals (IBI) were estimated from PPG signals. As IBI estimation is highly sensitive to motion or other types of noise, acceleration signals and PPG waveforms were used to automatically detect and discard unreliable IBI. AF was detected from windows of 20 consecutive IBI with 98.45±6.89% sensitivity and 99.13±1.79% specificity for 76.34±19.54% of the time. For the remaining time, no decision was taken due to the lack of reliable IBI. The results show that wrist PPG is suitable for long term monitoring and AF screening. In addition, this technique provides a more comfortable alternative to ECG devices.


  • Monitoring of heart rate and inter-beat intervals with wrist plethysmography in patients with atrial fibrillation

    by Jarkko Harju, Adrian Tarniceriu, Jakub Parak, Antti Vehkaoja, Arvi Yli-Hankala, Ilkka Korhonen

    Objective : Atrial fibrillation (AF) causes marked risk for patients, while silent fibrillation may remain unnoticed if not suspected and screened. Development of comfortable yet accurate beat-to-beat heart rate (HR) monitoring with good AF detection sensitivity would facilitate screening and improve treatment. The purpose of this study was to evaluate whether a wrist-worn photoplethysmography (PPG) device can be used to monitor beat-to-beat HR accurately during post-operative treatment in patients suffering from AF and whether wrist-PPG can be used to distinguish AF from sinus rhythm (SR). Approach : Twenty-nine patients (14 with AF, 15 with SR, mean age 71.5 years) with multiple comorbidities were monitored during routine post-operative treatment. The monitoring included standard ECG, finger PPG monitoring and a wrist-worn PPG monitor with green and infrared light sources. The HR from PPG sensors was compared against ECG-derived HR. Main results : The wrist PPG technology had very good HR and beat detection accuracy when using green light. For the SR group, the mean absolute error (MAE) for HR was 1.50 bpm, and for the inter-beat intervals (IBI), the MAE was 7.64 ms. For the AF group, the MAE for HR was 4.28 bpm and for IBI, the MAE was 14.67 ms. Accuracy for the infrared (IR) channel was worse. Finger PPG provided similar accuracy for HR and better accuracy for the IBI. AF detection sensitivity using green light was 99.0% and the specificity was 93.0%. Performance can be improved by discarding unreliable IBI periods. Significance : Results suggest that wrist PPG measurement allows accurate HR and beat-to-beat HR monitoring also in AF patients, and could be used for differentiating between SR and AF with very good sensitivity.


  • Atrial Fibrillation Detection from Wrist Photoplethysmography Data Using Artificial Neural Networks

    by Zeinab Rezaei Yousefi, Jakub Parak, Adrian Tarniceriu, Jarkko Harju, Arvi Yli-Hankala, Ilkka Korhonen, Antti Vehkaoja


  • Open Access

    Detection of beat-to-beat intervals from wrist photoplethysmography in patients with sinus rhythm and atrial fibrillation after surgery

    by Adrian Tarniceriu, Jarkko Harju, Antti Vehkaoja, Jakub Parak, Ricard Delgado-Gonzalo, Philippe Renevey, Arvi Yli-Hankala, Ilkka Korhonen

    Wrist photoplethysmography (PPG) allows unobtrusive monitoring of the heart rate (HR). PPG is affected by the capillary blood perfusion and the pumping function of the heart, which generally deteriorate with age and due to the presence of cardiac arrhythmia. The performance of wrist PPG in monitoring beat-to-beat HR in older patients with arrhythmia has not been reported earlier. We monitored PPG from wrist in 18 patients recovering from surgery in the post-anesthesia care unit, and evaluated the inter-beat interval (IBI) detection accuracy against ECG based R-to-R intervals (RRI). Nine subjects had sinus rhythm (SR, 68.0y ± 10.2y, 6 males) and nine subjects had atrial fibrillation (AF, 71.3y ± 7.8y, 4 males) during the recording. For the SR group, 99.44% of the beats were correctly identified, 2.39% extra beats were detected, and the mean absolute error (MAE) was 7.34 ms. For the AF group, 97.49% of the heartbeats were correctly identified, 2.26% extra beats were detected, and the MAE was 14.31 ms. IBI from the PPG were hence in close agreement with the ECG reference in both groups. The results suggest that wrist PPG provides a comfortable alternative to ECG during low motion and can be used for long-term monitoring and screening of AF episodes.


  • Learning a physical activity classifier for a low-power embedded wrist-located device

    by Ricard Delgado-Gonzalo, Philippe Renevey, Adrian Tarniceriu, Jakub Parak, Mattia Bertschi

    This article presents and evaluates a novel algorithm for learning a physical activity classifier for a low-power embedded wrist-located device. The overall system is designed for real-time execution and it is implemented in the commercial low-power System-on-Chips nRF51 and nRF52. Results were obtained using a database composed of 140 users containing more than 340 hours of labeled raw acceleration data. The final precision achieved for the most important classes, (Rest, Walk, and Run), was of 96%, 94%, and 99% and it generalizes to compound activities such as XC skiing or Housework. We conclude with a benchmarking of the system in terms of memory footprint and power consumption.


  • Open Access

    Estimating Heart Rate, Energy Expenditure, and Physical Performance With a Wrist Photoplethysmographic Device During Running

    by Jakub Parak, Maria Uuskoski, Jan Machek, Ilkka Korhonen

    Mobile and tablet apps, ubiquitous and pervasive computing, wearable computing and domotics for health.


  • Towards 24/7 continuous heart rate monitoring

    by Adrian Tarniceriu, Jakub Parak, Philippe Renevey, Marko Nurmi, Mattia Bertschi, Ricard Delgado-Gonzalo, Ilkka Korhonen

    Heart rate (HR) and HR variability (HRV) carry rich information about physical activity, mental and physical load, physiological status, and health of an individual. When combined with activity monitoring and personalized physiological modelling, HR/HRV monitoring may be used for monitoring of complex behaviors and impact of behaviors and external factors on the current physiological status of an individual. Optical HR monitoring (OHR) from wrist provides a comfortable and unobtrusive method for HR/HRV monitoring and is better adhered by users than traditional ECG electrodes or chest straps. However, OHR power consumption is significantly higher than that for ECG based methods due to the measurement principle based on optical illumination of the tissue. We developed an algorithmic approach to reduce power consumption of the OHR in 24/7 HR trending. We use continuous activity monitoring and a fast converging frequency domain algorithm to derive a reliable HR estimate in 7.1s (during outdoor sports, in average) to 10.0s (during daily life). The method allows >80% reduction in power consumption in 24/7 OHR monitoring when average HR monitoring is targeted, without significant reduction in tracking accuracy.


  • W2E-–Wellness Warehouse Engine for Semantic Interoperability of Consumer Health Data

    by Harri Honko, Vafa Andalibi, Timo Aaltonen, Jakub Parak, Mika Saaranen, Jari Viik, Ilkka Korhonen

    Novel health monitoring devices and applications allow consumers easy and ubiquitous ways to monitor their health status. However, technologies from different providers lack both technical and semantic interoperability and hence the resulting health data are often deeply tied to a specific service, which is limiting its reusability and utilization in different services. We have designed a Wellness Warehouse Engine (W2E) that bridges this gap and enables seamless exchange of data between different services. W2E provides interfaces to various data sources and makes data available via unified representational state transfer application programming interface to other services. Importantly, it includes Unifier--an engine that allows transforming input data into generic units reusable by other services, and Analyzer--an engine that allows advanced analysis of input data, such as combining different data sources into new output parameters. In this paper, we describe the architecture of W2E and demonstrate its applicability by using it for unifying data from four consumer activity trackers, using a test base of 20 subjects each carrying out three different tracking sessions. Finally, we discuss challenges of building a scalable Unifier engine for the ever-enlarging number of new devices.


  • Evaluation of the beat-to-beat detection accuracy of PulseOn wearable optical heart rate monitor

    by Jakub Parak, Adrian Tarniceriu, Philippe Renevey, Mattia Bertschi, Ricard Delgado-Gonzalo, Ilkka Korhonen

    Heart rate variability (HRV) provides significant information about the health status of an individual. Optical heart rate monitoring is a comfortable alternative to ECG based heart rate monitoring. However, most available optical heart rate monitoring devices do not supply beat-to-beat detection accuracy required by proper HRV analysis. We evaluate the beat-to-beat detection accuracy of a recent wrist-worn optical heart rate monitoring device, PulseOn (PO). Ten subjects (8 male and 2 female; 35.9±10.3 years old) participated in the study. HRV was recorded with PO and Firstbeat Bodyguard 2 (BG2) device, which was used as an ECG based reference. HRV was recorded during sleep. As compared to BG2, PO detected on average 99.57% of the heartbeats (0.43% of beats missed) and had 0.72% extra beat detection rate, with 5.94 ms mean absolute error (MAE) in beat-to-beat intervals (RRI) as compared to the ECG based RRI BG2. Mean RMSSD difference between PO and BG2 derived HRV was 3.1 ms. Therefore, PO provides an accurate method for long term HRV monitoring during sleep.