W.J. Bowers
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
W.J. Bowers has a total of 4 co-authors. Here are the top co-authors by number of shared publications:
- James W. Schmidt
National Institute of Standards and Technology (NIST), 100 Bureau Drive, Gaithersburg, MD 20899, United States of America
1 shared publication K. Jain
National Institute of Standards and Technology, Gaithersburg, MD 20899-0001
1 shared publication- Kiran Jain
National Solar Observatory
1 shared publication Y. Cen
National Institute of Standards and Technology, Gaithersburg, MD 20899-0001
1 shared publication
Latest Publications
- Open Access
- 1 March 2003
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 108 (2) , 135-45
A primary dead-weight tester for pressures (0.05-1.0) MPa
Recent advances in technology on two fronts, 1) the fabrication of large-diameter pistons and cylinders with good geometry, and 2) the ability to measure the dimensions of these components with high accuracy, have allowed dead-weight testers at the National Institute of Standards and Technology (NIST) to generate pressures that approach total relative uncertainties previously obtained only by manometers. This paper describes a 35 mm diameter piston/cylinder assembly (known within NIST as PG-39) that serves as a pressure standard in which both the piston and the cylinder have been accurately dimensioned by Physikalisch Technische Bundesanstalt (PTB). Both artifacts (piston and cylinder) appeared to be round within ±30 nm and straight within ±100 nm over a substantial fraction of their heights. Based on the diameters at 20 °C provided by PTB (±15 nm) and on the good geometry of the artifact, the relative uncertainties for the effective area were estimated to be about 2.2 × 10(-6) (1σ).
- 1 March 2003
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 108 (2) , 135-45
Open AccessA primary dead-weight tester for pressures (0.05-1.0) MPa
Recent advances in technology on two fronts, 1) the fabrication of large-diameter pistons and cylinders with good geometry, and 2) the ability to measure the dimensions of these components with high accuracy, have allowed dead-weight testers at the National Institute of Standards and Technology (NIST) to generate pressures that approach total relative uncertainties previously obtained only by manometers. This paper describes a 35 mm diameter piston/cylinder assembly (known within NIST as PG-39) that serves as a pressure standard in which both the piston and the cylinder have been accurately dimensioned by Physikalisch Technische Bundesanstalt (PTB). Both artifacts (piston and cylinder) appeared to be round within ±30 nm and straight within ±100 nm over a substantial fraction of their heights. Based on the diameters at 20 °C provided by PTB (±15 nm) and on the good geometry of the artifact, the relative uncertainties for the effective area were estimated to be about 2.2 × 10(-6) (1σ).
- 1 January 2003
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 108 (1) , 11-20
A double-primary dead-weight tester for pressures (35-175) kPa in gage mode
Primary pressure standards in the atmospheric pressure range are often established using mercury manometers. Less frequently, controlled-clearance dead-weight testers in which one component (normally the piston) has been dimensionally measured have also been used. Recent advances in technology on two fronts i) the fabrication of large-diameter pistons and cylinders with good geometry; and ii) the ability to measure the dimensions of these components, have allowed some dead-weight testers at NIST to approach total relative uncertainties (k = 2) in dimensionally-derived effective areas near 5 × 10(-6). This paper describes a single piston/cylinder assembly (NIST-PG201WC/WC) that serves as both a primary gage in which both piston and cylinder are measured dimensionally and a controlled-clearance primary gage (employing the Heydemann-Welch method). Thus it allows some previous assumptions about the modeling of dead-weight testers to be checked. For the gage described in this paper the piston/cylinder clearance obtained from the two analyses have relative differences of 4 × 10(-6) to 7 × 10(-6) over the pressure range 35 kPa to 175 kPa. Some implications of these results will be discussed. From the dimensional characterizations and auxiliary measurements we have determined that the effective area for this gauge at 20 °C is: [Formula: see text]where P is the system pressure and P J is a control pressure. The estimated relative uncertainty in effective area is 8.2 × 10(-6) +1.4 × 10(-11) P/Pa (k = 2). The temperature coefficient for the area was measured and found to be (9.06 ± 0.04) × 10(-6)/K. Thus using the gage at a reference temperature of 23 °C yields an effective area: [Formula: see text]with almost no increase in the uncertainty over that at 20 °C.
- 1 January 2003
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 108 (1) , 11-20
A double-primary dead-weight tester for pressures (35-175) kPa in gage mode
Primary pressure standards in the atmospheric pressure range are often established using mercury manometers. Less frequently, controlled-clearance dead-weight testers in which one component (normally the piston) has been dimensionally measured have also been used. Recent advances in technology on two fronts i) the fabrication of large-diameter pistons and cylinders with good geometry; and ii) the ability to measure the dimensions of these components, have allowed some dead-weight testers at NIST to approach total relative uncertainties (k = 2) in dimensionally-derived effective areas near 5 × 10(-6). This paper describes a single piston/cylinder assembly (NIST-PG201WC/WC) that serves as both a primary gage in which both piston and cylinder are measured dimensionally and a controlled-clearance primary gage (employing the Heydemann-Welch method). Thus it allows some previous assumptions about the modeling of dead-weight testers to be checked. For the gage described in this paper the piston/cylinder clearance obtained from the two analyses have relative differences of 4 × 10(-6) to 7 × 10(-6) over the pressure range 35 kPa to 175 kPa. Some implications of these results will be discussed. From the dimensional characterizations and auxiliary measurements we have determined that the effective area for this gauge at 20 °C is: [Formula: see text]where P is the system pressure and P J is a control pressure. The estimated relative uncertainty in effective area is 8.2 × 10(-6) +1.4 × 10(-11) P/Pa (k = 2). The temperature coefficient for the area was measured and found to be (9.06 ± 0.04) × 10(-6)/K. Thus using the gage at a reference temperature of 23 °C yields an effective area: [Formula: see text]with almost no increase in the uncertainty over that at 20 °C.