Masanori Ishii
Masanori Ishii has a total of 13 co-authors. Here are the top co-authors by number of shared publications:
- Yozo Shimada
National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Tsukuba, Japan
7 shared publications Koji Komiyama
3 shared publications- Moto Kinoshita
National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba, Japan
2 shared publications Satoru Kurokawa
National Institute of Advanced Industrial Science and Technology, National Metrology Institute of Japan, Tsukuba, Ibaraki, Japan
2 shared publications- Satoru Kurokawa
National Institute of Advanced Industrial Science and Technology (AIST)
2 shared publications Daisuke Nakajima
Japan Quality Assurance Organization, Japan
1 shared publication
Latest Publications
- 21 September 2020
- Published by Wiley in Microwave and Optical Technology Letters
- Vol. 63 (2) , 399-403
A feasibility study of a real‐time visualization method for electromagnetic fields
We present a feasibility study of a real‐time visualization method for electromagnetic fields. When a resonant laser is used to irradiate gaseous atoms enclosed in a glass cell, the atoms absorb the light and immediately radiate as fluorescence. If a resonant laser and electromagnetic field are used to irradiate the gaseous atoms simultaneously, fluorescence of the light is stronger. The difference between the strength values of fluorescence of light can be used for electromagnetic field visualization. We succeeded in demonstrating a real‐time visualization of electromagnetic fields radiated from a horn antenna and an aperture of a waveguide in free space.
- 21 September 2020
- Published by Wiley in Microwave and Optical Technology Letters
- Vol. 63 (2) , 399-403
A feasibility study of a real‐time visualization method for electromagnetic fields
We present a feasibility study of a real‐time visualization method for electromagnetic fields. When a resonant laser is used to irradiate gaseous atoms enclosed in a glass cell, the atoms absorb the light and immediately radiate as fluorescence. If a resonant laser and electromagnetic field are used to irradiate the gaseous atoms simultaneously, fluorescence of the light is stronger. The difference between the strength values of fluorescence of light can be used for electromagnetic field visualization. We succeeded in demonstrating a real‐time visualization of electromagnetic fields radiated from a horn antenna and an aperture of a waveguide in free space.
- 1 August 2017
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 84-89
Digital comb-generator for validating sub-30 MHz radiated emission test system using double-gapped shielded loop antenna
Recently, the needs for the emission measurement of electronic devices are increasing. Especially, in association with the development of RFID system, electromagnetic cooker, and the wireless power transmission for automobile, the radiated emission measurement in the frequency range from 9 kHz to 30 MHz become very important. Comb-generators as a reference signal source are very useful equipment for validating the radiated emission test system. Conventional comb-generators periodically generate pulse waveforms with a certain period, T. As a result, comb-like spectrum with the frequency-spacing of 1/T are generated from the conventional comb-generators. The period of pulse generation and the output level of comb-generators have inverse relationship because longer period decrease the waveform energy per second. In addition, the frequency spectrum of the conventional comb-generator and the pulse waveform are Fourier transformation pairs. Therefore, the fine control of the spectrum of the comb-generator is quite difficult. Additionally, if the frequency-spacing of the generator is set to 1 kHz, the frequency spectrum is too dense around 30 MHz and the output level of the comb-generator are too low due to the inverse relationship of the period and the output level. On the other hand, if the frequency-spacing of the generator is set to 1 MHz, the spectrum of less than 1 MHz cannot be generated. To solve these problems, a digital comb-generator using FPGA-based arbitrary waveform generator and double-gapped shielded loop antenna is proposed in this paper. This digital comb-generator can cover the frequency range from 150 kHz to 30 MHz, which correspond to CISPR Band B. The proposed comb-generator can also generate almost equally-spaced comb-like spectrum in the logarithmic frequency scale. For emission test in the frequency bellow 30 MHz, the log-scale frequency scale is frequently used. Therefore, the log-scale comb-generator is very useful for full-band measurement in the frequency range from 9 kHz to 30 MHz. Moreover, by the use of double-gapped shielded loop antenna, the proposed comb-generator can generate both magnetic-field and electric-field by controlling the phase of waveform. The generated magnetic field strength from the proposed comb generator is about 44 dBuA/m to 54 dBuA/m. Additionally, the standard deviation in 3-hour measurement of the output spectrum is less than 0.05 dB in the frequency less than 30 MHz. Therefore, the output spectrum of the proposed comb-generator is quite stable. From these results, the proposed comb-generator can be used as a reference signal source for validating sub-30 MHz emission test systems.
- 1 August 2017
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 84-89
Digital comb-generator for validating sub-30 MHz radiated emission test system using double-gapped shielded loop antenna
Recently, the needs for the emission measurement of electronic devices are increasing. Especially, in association with the development of RFID system, electromagnetic cooker, and the wireless power transmission for automobile, the radiated emission measurement in the frequency range from 9 kHz to 30 MHz become very important. Comb-generators as a reference signal source are very useful equipment for validating the radiated emission test system. Conventional comb-generators periodically generate pulse waveforms with a certain period, T. As a result, comb-like spectrum with the frequency-spacing of 1/T are generated from the conventional comb-generators. The period of pulse generation and the output level of comb-generators have inverse relationship because longer period decrease the waveform energy per second. In addition, the frequency spectrum of the conventional comb-generator and the pulse waveform are Fourier transformation pairs. Therefore, the fine control of the spectrum of the comb-generator is quite difficult. Additionally, if the frequency-spacing of the generator is set to 1 kHz, the frequency spectrum is too dense around 30 MHz and the output level of the comb-generator are too low due to the inverse relationship of the period and the output level. On the other hand, if the frequency-spacing of the generator is set to 1 MHz, the spectrum of less than 1 MHz cannot be generated. To solve these problems, a digital comb-generator using FPGA-based arbitrary waveform generator and double-gapped shielded loop antenna is proposed in this paper. This digital comb-generator can cover the frequency range from 150 kHz to 30 MHz, which correspond to CISPR Band B. The proposed comb-generator can also generate almost equally-spaced comb-like spectrum in the logarithmic frequency scale. For emission test in the frequency bellow 30 MHz, the log-scale frequency scale is frequently used. Therefore, the log-scale comb-generator is very useful for full-band measurement in the frequency range from 9 kHz to 30 MHz. Moreover, by the use of double-gapped shielded loop antenna, the proposed comb-generator can generate both magnetic-field and electric-field by controlling the phase of waveform. The generated magnetic field strength from the proposed comb generator is about 44 dBuA/m to 54 dBuA/m. Additionally, the standard deviation in 3-hour measurement of the output spectrum is less than 0.05 dB in the frequency less than 30 MHz. Therefore, the output spectrum of the proposed comb-generator is quite stable. From these results, the proposed comb-generator can be used as a reference signal source for validating sub-30 MHz emission test systems.
- 20 February 2017
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 66 (6) , 1592-1597
Electromagnetic Field Sensor Based on Atomic Candle
A new electromagnetic sensor using the atomic Rabi frequency is presented. Gaseous cesium-133 atoms enclosed in a compact glass cell work as an electromagnetic sensor by detecting the Rabi frequency. In this paper, we measured the electromagnetic field radiated from a standard gain horn antenna using the new sensor. This result was compared with a numerical simulation based on the method of moments. Both results were in good agreement. Moreover, a hypothesized reduction in electromagnetic perturbation caused by the glass cell was investigated using a numerical simulation.
- 20 February 2017
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 66 (6) , 1592-1597
Electromagnetic Field Sensor Based on Atomic Candle
A new electromagnetic sensor using the atomic Rabi frequency is presented. Gaseous cesium-133 atoms enclosed in a compact glass cell work as an electromagnetic sensor by detecting the Rabi frequency. In this paper, we measured the electromagnetic field radiated from a standard gain horn antenna using the new sensor. This result was compared with a numerical simulation based on the method of moments. Both results were in good agreement. Moreover, a hypothesized reduction in electromagnetic perturbation caused by the glass cell was investigated using a numerical simulation.
- 1 September 2014
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 23250356,p. 934-939
A study on the characteristics of semi-anechoic chambers below 30 MHz
Several methods for evaluating test sites for EMI measurements above 30 MHz have been regulated by the CISPR, ANSI etc. Nowadays the demand for EMI measurements below 30 MHz is increasing; however, the methods to evaluate the test sites for such EMI measurements in the low frequency have not yet been regulated. In general, a loop antenna is used for measuring magnetic fields below 30 MHz, whereas a monopole antenna is used for measuring electric fields in the frequency band. Recently, we measured the characteristics of several semi-anechoic chambers (SACs) using monopole antennas and loop antennas. In this paper, we discuss and report a study on the characteristics of two types of SACs through measurement and simulation.
- 1 September 2014
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 23250356,p. 934-939
A study on the characteristics of semi-anechoic chambers below 30 MHz
Several methods for evaluating test sites for EMI measurements above 30 MHz have been regulated by the CISPR, ANSI etc. Nowadays the demand for EMI measurements below 30 MHz is increasing; however, the methods to evaluate the test sites for such EMI measurements in the low frequency have not yet been regulated. In general, a loop antenna is used for measuring magnetic fields below 30 MHz, whereas a monopole antenna is used for measuring electric fields in the frequency band. Recently, we measured the characteristics of several semi-anechoic chambers (SACs) using monopole antennas and loop antennas. In this paper, we discuss and report a study on the characteristics of two types of SACs through measurement and simulation.
- 5 February 2013
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 62 (6) , 1443-1449
Improvement of Formula and Uncertainty of the Reference Magnetic Field for AC Magnetometer Calibration
In this paper, an ac magnetic field standard system for ac magnetometer calibration is described. The target field strength and frequency range are 7.162 A/m (9 μT ) and 50-60 Hz, respectively. The system to generate a reference magnetic field is based on the Helmholtz coil formula. However, this well-known formula is precise only for dc magnetic fields. Therefore, we improve the formula for ac magnetic fields. The uncertainties of the reference magnetic field and magnetometer calibration are also improved. Validation is achieved by using two calibration systems based on different principles.
- 5 February 2013
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 62 (6) , 1443-1449
Improvement of Formula and Uncertainty of the Reference Magnetic Field for AC Magnetometer Calibration
In this paper, an ac magnetic field standard system for ac magnetometer calibration is described. The target field strength and frequency range are 7.162 A/m (9 μT ) and 50-60 Hz, respectively. The system to generate a reference magnetic field is based on the Helmholtz coil formula. However, this well-known formula is precise only for dc magnetic fields. Therefore, we improve the formula for ac magnetic fields. The uncertainties of the reference magnetic field and magnetometer calibration are also improved. Validation is achieved by using two calibration systems based on different principles.
- 1 August 2011
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 101-106
Comparison between three-antenna method and equivalent capacitance substitution method for calibrating electrically short monopole antenna
In low-frequency bands, monopole antennas are generally used for measuring the electromagnetic interference (EMI) and estimating the electric field strength. The antenna factor is an important and widely studied characteristic. Thus far, there have been a number of measurement methods proposed for electrically short monopole antennas. A case in point is the equivalent capacitance substitution method, which is commonly used for monopole antenna measurements. On the other hand, we have proposed a near-field three-antenna method for electrically short monopole antennas. These two absolute measurement methods are of quite different origins. In light of this fact, a comparison is made between the two measurement methods by means of experiments and simulations, and the difference observed between the antenna factors is discussed.
- 1 August 2011
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 101-106
Comparison between three-antenna method and equivalent capacitance substitution method for calibrating electrically short monopole antenna
In low-frequency bands, monopole antennas are generally used for measuring the electromagnetic interference (EMI) and estimating the electric field strength. The antenna factor is an important and widely studied characteristic. Thus far, there have been a number of measurement methods proposed for electrically short monopole antennas. A case in point is the equivalent capacitance substitution method, which is commonly used for monopole antenna measurements. On the other hand, we have proposed a near-field three-antenna method for electrically short monopole antennas. These two absolute measurement methods are of quite different origins. In light of this fact, a comparison is made between the two measurement methods by means of experiments and simulations, and the difference observed between the antenna factors is discussed.
- 1 July 2010
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 2158110X,p. 56-61
A reference antenna method for non-resonant electrically short monopole antennas
In order to measure the electric field strength in low-frequency bands, monopole antennas are generally used. When carrying out EMI measurements, the antenna factor is an important and well-known characteristic. Generally, absolute calibration methods are commonly used for monopole antenna calibrations. However, establishing and keeping the traceability to standard quantities in absolute calibration methods are time consuming. In this paper, a new reference antenna method for non-resonant electrically short monopole antennas is proposed and investigated through a numerical simulation. Since the target frequency band is less than 30 MHz, the monopole antennas work as electrically short monopole antennas. The measurements are performed under the near-field condition. The results show that the proposed reference antenna method is available and effective in obtaining the antenna factor of a non-resonant electrically short monopole antenna.
- 1 July 2010
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 2158110X,p. 56-61
A reference antenna method for non-resonant electrically short monopole antennas
In order to measure the electric field strength in low-frequency bands, monopole antennas are generally used. When carrying out EMI measurements, the antenna factor is an important and well-known characteristic. Generally, absolute calibration methods are commonly used for monopole antenna calibrations. However, establishing and keeping the traceability to standard quantities in absolute calibration methods are time consuming. In this paper, a new reference antenna method for non-resonant electrically short monopole antennas is proposed and investigated through a numerical simulation. Since the target frequency band is less than 30 MHz, the monopole antennas work as electrically short monopole antennas. The measurements are performed under the near-field condition. The results show that the proposed reference antenna method is available and effective in obtaining the antenna factor of a non-resonant electrically short monopole antenna.
- 1 June 2010
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 05891485,p. 704-705
Measurement of electrically short monopole antenna by three-antenna method
In low frequency bands, monopole antenna and its antenna factor are used for electric field strength measurement. Three-antenna method is well-known method as one of the antenna calibration methods. However, standard field methods are widely employed in the monopole antenna calibration at national metrology institutes. Recently, we have proposed a three-antenna method for electrically short monopole antenna and investigated the method by a numerical simulation. In this paper, we perform measurements for actual electrically short monopole antennas and investigate the proposed method by experiments.
- 1 June 2010
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 05891485,p. 704-705
Measurement of electrically short monopole antenna by three-antenna method
In low frequency bands, monopole antenna and its antenna factor are used for electric field strength measurement. Three-antenna method is well-known method as one of the antenna calibration methods. However, standard field methods are widely employed in the monopole antenna calibration at national metrology institutes. Recently, we have proposed a three-antenna method for electrically short monopole antenna and investigated the method by a numerical simulation. In this paper, we perform measurements for actual electrically short monopole antennas and investigate the proposed method by experiments.
- 1 August 2009
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 324-327
A near field 3-antenna method for short monopole antennas in low frequency bands
There are some measurement methods proposed for an electrically short monopole antenna. In this paper, a new three-antenna method for electrically short monopole antennas is proposed and investigated by a numerical simulation. The target of the frequency bands is less than 30 MHz. The results show that the proposed measurement method is available and effective to obtain the antenna factor of an electrically short monopole antenna.
- 1 August 2009
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 324-327
A near field 3-antenna method for short monopole antennas in low frequency bands
There are some measurement methods proposed for an electrically short monopole antenna. In this paper, a new three-antenna method for electrically short monopole antennas is proposed and investigated by a numerical simulation. The target of the frequency bands is less than 30 MHz. The results show that the proposed measurement method is available and effective to obtain the antenna factor of an electrically short monopole antenna.
- 20 January 2009
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 58 (4) , 1097-1103
Reference Calibration Methods for Small Circular Loop Antenna in Low-Frequency Band
In this paper, we propose two reference calibration methods (RCMs) for small circular loop antennas. Generally, the standard field method is widely employed in loop antenna calibration at many national metrology institutes and calibration laboratories. In this method, a thermocouple is employed as a current sensor on an antenna element of a transmitting loop antenna to calculate the magnetic field strength at an antenna under calibration. This method has certain disadvantages: The standard transmitting antenna is complicated, cumbersome, and fragile. On the other hand, a calibration method is necessary to be reasonably traceable to the standard of the magnetic antenna factor. Our proposed methods can establish and keep traceability to it. We also compare the two proposed RCMs and show that they are efficient in calibrating the loop antennas.
- 20 January 2009
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 58 (4) , 1097-1103
Reference Calibration Methods for Small Circular Loop Antenna in Low-Frequency Band
In this paper, we propose two reference calibration methods (RCMs) for small circular loop antennas. Generally, the standard field method is widely employed in loop antenna calibration at many national metrology institutes and calibration laboratories. In this method, a thermocouple is employed as a current sensor on an antenna element of a transmitting loop antenna to calculate the magnetic field strength at an antenna under calibration. This method has certain disadvantages: The standard transmitting antenna is complicated, cumbersome, and fragile. On the other hand, a calibration method is necessary to be reasonably traceable to the standard of the magnetic antenna factor. Our proposed methods can establish and keep traceability to it. We also compare the two proposed RCMs and show that they are efficient in calibrating the loop antennas.