Joel Rutkowski
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Joel Rutkowski has a total of 28 co-authors. Here are the top co-authors by number of shared publications:
- Dave Harber
Laboratory for Atmospheric and Space Physics University of Colorado Boulder Boulder CO USA
3 shared publications - Erik Richard
Laboratory for Atmospheric and Space Physics, Boulder, CO 80303, United States of America
3 shared publications - Karl Heuerman
Laboratory for Atmospheric and Space Physics, Boulder, CO 80303, United States of America
3 shared publications - Nathan A. Tomlin
National Institute of Standards and Technology, Boulder, CO 80305, United States of America
3 shared publications - Cameron Straatsma
Laboratory for Atmospheric and Space Physics, Boulder, CO 80303, United States of America
2 shared publications - Christopher S. Yung
Applied Physics Division, National Institute of Standards and Technology, Boulder, USA
2 shared publications
Latest Publications
- Open Access
- 19 January 2026
- Published by IOP Publishing in Metrologia
- Vol. 63 (1)
Applying Heat Transport Modelling to Reduce Measurement Uncertainty of the LASP Reference Standard NACR8
Heat transport modelling of a planar absolute detector with vertically aligned carbon nanotube absorbers has been applied to the newly developed ambient-temperature electrical substitution radiometer (ESR), NACR8. Integrated into the Laboratory for Atmospheric and Space Physics (LASP) Total-solar-irradiance Radiometer Facility (TRF) NACR8 operates as an absolute optical power reference standard supporting validation of satellite based solar irradiance measurements, including those from the TSIS 2 TIM instrument. The modelling enables estimation and correction for electro-optical non-equivalence and its associated uncertainty, providing a generalised method applicable to planar thermal detectors. This approach represents an advancement in the understanding and correction of radiometric non-equivalence in optical radiation metrology. Traceability and performance of NACR8, incorporating a correction factor of 0.999562 with a standard uncertainty (u) = 0.000183 were confirmed through comparison with the National Institute of Standards and Technology (NIST) Primary Optical Watt Radiometer (POWR) at 532.12 nm using a silicon photodiode light trapping transfer standard. Further comparison with NACR5, an established LASP optical power standard, showed agreement within stated uncertainties demonstrating the effectiveness of the heat transport modelling approach for addressing electro-optical non-equivalence in planar thermal detectors.
- 19 January 2026
- Published by IOP Publishing in Metrologia
- Vol. 63 (1)
Open AccessApplying Heat Transport Modelling to Reduce Measurement Uncertainty of the LASP Reference Standard NACR8
Heat transport modelling of a planar absolute detector with vertically aligned carbon nanotube absorbers has been applied to the newly developed ambient-temperature electrical substitution radiometer (ESR), NACR8. Integrated into the Laboratory for Atmospheric and Space Physics (LASP) Total-solar-irradiance Radiometer Facility (TRF) NACR8 operates as an absolute optical power reference standard supporting validation of satellite based solar irradiance measurements, including those from the TSIS 2 TIM instrument. The modelling enables estimation and correction for electro-optical non-equivalence and its associated uncertainty, providing a generalised method applicable to planar thermal detectors. This approach represents an advancement in the understanding and correction of radiometric non-equivalence in optical radiation metrology. Traceability and performance of NACR8, incorporating a correction factor of 0.999562 with a standard uncertainty (u) = 0.000183 were confirmed through comparison with the National Institute of Standards and Technology (NIST) Primary Optical Watt Radiometer (POWR) at 532.12 nm using a silicon photodiode light trapping transfer standard. Further comparison with NACR5, an established LASP optical power standard, showed agreement within stated uncertainties demonstrating the effectiveness of the heat transport modelling approach for addressing electro-optical non-equivalence in planar thermal detectors.
- Open Access
- 16 August 2022
- Published by IOP Publishing in Metrologia
- Vol. 59 (6) , 065006
Decadal validation of the LASP TRF cryogenic radiometer by NIST, and establishment of a replacement room temperature standard*
We present the results of a recent, extensive measurement campaign validating the traceability of the solar irradiance record and Earth radiation budget data. The campaign also established future traceability, thus ensuring confidence in the continuing climate-data record. The total solar irradiance radiometer facility (TRF) at the Laboratory for Atmospheric and Space Physics (LASP) Boulder, uses a liquid helium cooled cryogenic radiometer as the reference standard for the validation of spaceflight total solar irradiance (TSI) instrumentation. In 2008 the radiometer was directly compared to the National Institute of Standards and Technology (NIST) Primary Optical Watt Radiometer (POWR) at a wavelength of 532.12 nm. At TSI power levels, a correction factor of 1.000 306 with an associated standard uncertainty (u) of 0.000 098, was reported for the TRF radiometer scale when using external voltage measurement electronics, and not correcting for cavity heating non-equivalence or cavity absorptance. The TRF radiometer has recently been revalidated at LASP using a POWR calibrated silicon photodiode trap transfer standard named TT4. We report a correction factor of 0.999 787, u = 0.000 285 to align the TRF radiometer scale with the current NIST POWR scale. A new room temperature reference standard radiometer was established. It measured 133 parts per million (ppm) higher than POWR using the same silicon transfer standard as above, and in a separate direct measurement, 168 ppm lower than the TRF radiometer shuttered at 400 s full shutter cycle. The difference agrees within stated uncertainties. A correction of 0.999 867, u = 0.000 247 will align the new radiometer scale with the NIST radiant power scale of POWR.
- 16 August 2022
- Published by IOP Publishing in Metrologia
- Vol. 59 (6) , 065006
Open AccessDecadal validation of the LASP TRF cryogenic radiometer by NIST, and establishment of a replacement room temperature standard*
We present the results of a recent, extensive measurement campaign validating the traceability of the solar irradiance record and Earth radiation budget data. The campaign also established future traceability, thus ensuring confidence in the continuing climate-data record. The total solar irradiance radiometer facility (TRF) at the Laboratory for Atmospheric and Space Physics (LASP) Boulder, uses a liquid helium cooled cryogenic radiometer as the reference standard for the validation of spaceflight total solar irradiance (TSI) instrumentation. In 2008 the radiometer was directly compared to the National Institute of Standards and Technology (NIST) Primary Optical Watt Radiometer (POWR) at a wavelength of 532.12 nm. At TSI power levels, a correction factor of 1.000 306 with an associated standard uncertainty (u) of 0.000 098, was reported for the TRF radiometer scale when using external voltage measurement electronics, and not correcting for cavity heating non-equivalence or cavity absorptance. The TRF radiometer has recently been revalidated at LASP using a POWR calibrated silicon photodiode trap transfer standard named TT4. We report a correction factor of 0.999 787, u = 0.000 285 to align the TRF radiometer scale with the current NIST POWR scale. A new room temperature reference standard radiometer was established. It measured 133 parts per million (ppm) higher than POWR using the same silicon transfer standard as above, and in a separate direct measurement, 168 ppm lower than the TRF radiometer shuttered at 400 s full shutter cycle. The difference agrees within stated uncertainties. A correction of 0.999 867, u = 0.000 247 will align the new radiometer scale with the NIST radiant power scale of POWR.
- Open Access
- 1 May 2020
- Published by AIP Publishing in AIP Advances
- Vol. 10 (5) , 055010
Overview of microfabricated bolometers with vertically aligned carbon nanotube absorbers
Multi-wall vertically aligned carbon nanotubes (VACNTs) are nearly ideal absorbers due to their exceptionally low reflectance over a broad wavelength range. Integrating VACNTs as bolometer absorbers, however, can be difficult due to their high growth temperature and fragile nature. Despite these challenges, we have microfabricated many different types of VACNT bolometers, ranging from cryogenic optical power primary standards to room temperature satellite-based solar irradiance monitors and broadband infrared microbolometers. Advantages our VACNT bolometers provide over the bolometers they replace vary by application, but can be reduced size and time constant, increased absorption, and/or microfabrication instead of hand assembly. Depending on the application and operating conditions, our VACNT bolometers are designed with a variety of thermistors and weak thermal links. The thermistors used include commercial surface mount chips, superconducting transition-edge sensors, and vanadium oxide (VOx). Weak thermal links include silicon nitride (SiNx) membranes, Si bridges, and laser-cut polyimide. We summarize a wide variety of microfabricated bolometers with VACNT absorbers that measure optical power levels spanning over seven orders of magnitude.
- 1 May 2020
- Published by AIP Publishing in AIP Advances
- Vol. 10 (5) , 055010
Open AccessOverview of microfabricated bolometers with vertically aligned carbon nanotube absorbers
Multi-wall vertically aligned carbon nanotubes (VACNTs) are nearly ideal absorbers due to their exceptionally low reflectance over a broad wavelength range. Integrating VACNTs as bolometer absorbers, however, can be difficult due to their high growth temperature and fragile nature. Despite these challenges, we have microfabricated many different types of VACNT bolometers, ranging from cryogenic optical power primary standards to room temperature satellite-based solar irradiance monitors and broadband infrared microbolometers. Advantages our VACNT bolometers provide over the bolometers they replace vary by application, but can be reduced size and time constant, increased absorption, and/or microfabrication instead of hand assembly. Depending on the application and operating conditions, our VACNT bolometers are designed with a variety of thermistors and weak thermal links. The thermistors used include commercial surface mount chips, superconducting transition-edge sensors, and vanadium oxide (VOx). Weak thermal links include silicon nitride (SiNx) membranes, Si bridges, and laser-cut polyimide. We summarize a wide variety of microfabricated bolometers with VACNT absorbers that measure optical power levels spanning over seven orders of magnitude.