Peter Matkulcik
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Previous Organizations / Institutes
The organizations / institutes from the scholar's other articles in the past ten years
Peter Matkulcik has a total of 10 co-authors. Here are the top co-authors by number of shared publications:
- Dalibor Cervinka
Faculty of Electrical Engineering and Communication, Department of Power Electrical and Electronic Engineering, Brno University of Technology
2 shared publications - Shraga Nahum Goldberg
Division of Image-Guided Therapy, Department of Radiology, Hadassah Hebrew University Medical Center, Jerusalem, Israel.
2 shared publications - Tomas Rohan
Faculty of Medicine Masaryk University Brno Czechia
2 shared publications - Tomáš Andrašina
Department of Radiology and Nuclear Medicine Faculty of Medicine University Hospital Brno Masaryk University Brno Czech Republic
2 shared publications - Veronika Novotna
FEEC BUT
2 shared publications I. Svobodova
Department of Pathology, St. Anne's University Hospital Brno, Brno, Czech Republic
1 shared publication
Latest Publications
- Open Access
- 2 July 2025
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 42 (1) , 2520362
Mathematical modeling of the effect of stent construction during endoluminal IRE for recanalization of an occluded metal stent
Intraluminal irreversible electroporation (IRE) can be used for recanalizing occluded metal stents. However, optimal IRE parameters for consistent effects across different stent designs remain unclear. The aim of this study was to simulate the process of stent recanalization in silico by employing finite element analysis. A virtual model of an occluded biliary stent with an experimental 3-electrode IRE catheter was developed. Electric field distribution, temperature changes, and potential ablation volumes were simulated across various parameters: IRE voltage (300 - 1300 V), stent wire width (0.1 - 0.5 mm) and stent mesh size (0.7 - 5.58 mm). Simulations incorporated five representative stent types commonly used in clinical practice. 685 unique simulations were conducted, analyzing 1162 unique values. Higher voltages generally led to larger ablation zones and increased temperatures. Thinner stent wires and larger mesh sizes also increased the extent of ablation zone. While in-stent ablation was largely independent of stent design, out-of-stent ablation was significantly impacted by mesh size and tissue thickness between the stent and irreversible electroporation electrodes. Voltages above 1000 V produced significant thermal effects, with substantial volumes of tissue heated above 50 °C. Specific stent designs exhibited variations in maximum temperature (72.1 - 83.1 °C) and ablation volume (8.7 - 14.7 mm3). Tailored IRE protocols for different stent designs are required due to differences in in- and out-stent ablation volumes. High voltages (>1000 V) induce both thermal and nonthermal ablation mechanisms.
- 2 July 2025
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 42 (1) , 2520362
Open AccessMathematical modeling of the effect of stent construction during endoluminal IRE for recanalization of an occluded metal stent
Intraluminal irreversible electroporation (IRE) can be used for recanalizing occluded metal stents. However, optimal IRE parameters for consistent effects across different stent designs remain unclear. The aim of this study was to simulate the process of stent recanalization in silico by employing finite element analysis. A virtual model of an occluded biliary stent with an experimental 3-electrode IRE catheter was developed. Electric field distribution, temperature changes, and potential ablation volumes were simulated across various parameters: IRE voltage (300 - 1300 V), stent wire width (0.1 - 0.5 mm) and stent mesh size (0.7 - 5.58 mm). Simulations incorporated five representative stent types commonly used in clinical practice. 685 unique simulations were conducted, analyzing 1162 unique values. Higher voltages generally led to larger ablation zones and increased temperatures. Thinner stent wires and larger mesh sizes also increased the extent of ablation zone. While in-stent ablation was largely independent of stent design, out-of-stent ablation was significantly impacted by mesh size and tissue thickness between the stent and irreversible electroporation electrodes. Voltages above 1000 V produced significant thermal effects, with substantial volumes of tissue heated above 50 °C. Specific stent designs exhibited variations in maximum temperature (72.1 - 83.1 °C) and ablation volume (8.7 - 14.7 mm3). Tailored IRE protocols for different stent designs are required due to differences in in- and out-stent ablation volumes. High voltages (>1000 V) induce both thermal and nonthermal ablation mechanisms.
- Open Access
- 1 January 2021
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 38 (1) , 393-401
Experimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter
To demonstrate the feasibility of irreversible electroporation (IRE) for treating biliary metal stent occlusion in an experimental liver model.IRE was performed using an expandable tubular IRE-cath...
- 1 January 2021
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 38 (1) , 393-401
Open AccessExperimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter
To demonstrate the feasibility of irreversible electroporation (IRE) for treating biliary metal stent occlusion in an experimental liver model.IRE was performed using an expandable tubular IRE-cath...