HNHamid Naraghi

Hamid Naraghi

*This is an algorithmically generated author record based on publicly available information.

Current Organizations / Institutes

The organizations/institutes from the scholar's latest article


Hamid Naraghi has a total of 6 co-authors. Here are the top co-authors by number of shared publications:


Latest Publications

  • Early Results after Aortic Valve Replacement Using Last Generation Bioprosthetic Aortic Valve

    by Dritan Useini, Markus Schlömicher, Peter Haldenwang, Hamid Naraghi, Vadim Moustafine, Matthias Bechtel, Justus Strauch

    Introduction: Few data are available about the newest generation surgical bioprosthesis. We aimed to evaluate early clinical and hemodynamic outcomes after using the INSPIRIS RESILIA aortic valve (Edwards Lifesciences, Irvine, California, USA). Methods: Between July 2018 and April 2021, 80 patients underwent aortic valve replacement receiving the INSPIRIS RESILIA aortic valve at our institution. Primary outcomes were the composite of early mortality, stroke, and myocardial infarction. Secondary outcomes were hemodynamic performances of the valve, paravalvular leakage, and new pacemaker implantation. Results: The mean age of the study population was 60.6 ± 8.3 years. The mean Society of Thoracic Surgery-Predicted Risk of Mortality score was 2.9 ± 1.7%. In 43.7% of the patients, concomitant surgery was performed. The in-hospital mortality, all-stroke, and myocardial infarction rates were 2.5%, 1.2%, and 1.2%, respectively. No valve was explanted and no redo was performed. The mean postoperative trans-prosthetic gradient at discharge was 10.2 ± 4.1 mm Hg. There was no need for new pacemaker implantation. We registered only two cases with minimal (trace) paravalvular leakage. Conclusion: The use of the INSPIRIS RESILIA aortic valve in a young, low-risk population is safe and associated with very good early clinical and hemodynamic outcomes.