Shizuka No
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Shizuka No has a total of 10 co-authors. Here are the top co-authors by number of shared publications:
- Hiroyuki Isayama
Juntendo University School of Medicine Graduate School of Medicine, Tokyo, Japan
1 shared publication - Kouta Inoue
Medical Laboratory, Research & Development Center, Zeon Corporation, Toyama, Japan
1 shared publication - Naoki Sasahira
Department of Hepato-Biliary-Pancreatic Medicine, Cancer Institute Hospital, Japanese Foundation for Cancer Research, Tokyo 135-8550, Japan
1 shared publication - Rei Ishibashi
Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo 113-8655, Japan
1 shared publication - Shuntaro Yoshida
Waseda University, Shinjuku, Tokyo, Japan
1 shared publication - Takashi Sasaki
Department of Hepato-Biliary-Pancreatic Medicine, Cancer Institute Hospital, Japanese Foundation for Cancer Research, Tokyo 135-8550, Japan
1 shared publication
Latest Publications
- Open Access
- 30 September 2023
- Published by The Korean Society of Gastrointestinal Endoscopy in Clinical Endoscopy
- Vol. 56 (5) , 633-649
Evaluation of the mechanical properties of current biliary self-expandable metallic stents: axial and radial force, and axial force zero border
Background/Aims: Mechanical properties (MPs) and axial and radial force (AF and RF) may influence the efficacy and complications of self-expandable metallic stent (SEMS) placement. We measured the MPs of various SEMSs and examined their influence on the SEMS clinical ability.Methods: We evaluated the MPs of 29 types of 10-mm SEMSs. RF was measured using a conventional measurement device. AF was measured using the conventional and new methods, and the correlation between the methods was evaluated.Results: A high correlation in AFs was observed, as measured by the new and conventional manual methods. AF and RF scatterplots divided the SEMSs into three subgroups according to structure: hook-and-cross-type (low AF and RF), cross-type (high AF and low RF), and laser-cut-type (intermediate AF and high RF). The hook-and-cross-type had the largest axial force zero border (>20°), followed by the laser-cut and cross types.Conclusions: MPs were related to stent structure. Hook-and-cross-type SEMSs had a low AF and high axial force zero border and were considered safest because they caused minimal stress on the biliary wall. However, the increase in RF must be overcome.
- 30 September 2023
- Published by The Korean Society of Gastrointestinal Endoscopy in Clinical Endoscopy
- Vol. 56 (5) , 633-649
Open AccessEvaluation of the mechanical properties of current biliary self-expandable metallic stents: axial and radial force, and axial force zero border
Background/Aims: Mechanical properties (MPs) and axial and radial force (AF and RF) may influence the efficacy and complications of self-expandable metallic stent (SEMS) placement. We measured the MPs of various SEMSs and examined their influence on the SEMS clinical ability.Methods: We evaluated the MPs of 29 types of 10-mm SEMSs. RF was measured using a conventional measurement device. AF was measured using the conventional and new methods, and the correlation between the methods was evaluated.Results: A high correlation in AFs was observed, as measured by the new and conventional manual methods. AF and RF scatterplots divided the SEMSs into three subgroups according to structure: hook-and-cross-type (low AF and RF), cross-type (high AF and low RF), and laser-cut-type (intermediate AF and high RF). The hook-and-cross-type had the largest axial force zero border (>20°), followed by the laser-cut and cross types.Conclusions: MPs were related to stent structure. Hook-and-cross-type SEMSs had a low AF and high axial force zero border and were considered safest because they caused minimal stress on the biliary wall. However, the increase in RF must be overcome.