Evaluation of the mechanical properties of current biliary self-expandable metallic stents: axial and radial force, and axial force zero border
Authors:
- Wataru YamagataWYWataru Yamagata
- Toshio FujisawaTFToshio Fujisawa
- Takashi SasakiTSTakashi Sasaki
- Rei IshibashiRIRei Ishibashi
- Tomotaka SaitoTSTomotaka Saito
- Shuntaro YoshidaSYShuntaro Yoshida
- Shizuka NoSNShizuka No
- Kouta InoueKIKouta Inoue
- Yousuke NakaiYNYousuke Nakai
- Naoki SasahiraNSNaoki Sasahira
- Hiroyuki IsayamaHiroyuki Isayama
Open Access
- 30 September 2023
- journal article
- research article
- Published by The Korean Society of Gastrointestinal Endoscopy in Clinical Endoscopy
- Vol. 56 (5) , 633-649
Abstract
Background/Aims: Mechanical properties (MPs) and axial and radial force (AF and RF) may influence the efficacy and complications of self-expandable metallic stent (SEMS) placement. We measured the MPs of various SEMSs and examined their influence on the SEMS clinical ability.Methods: We evaluated the MPs of 29 types of 10-mm SEMSs. RF was measured using a conventional measurement device. AF was measured using the conventional and new methods, and the correlation between the methods was evaluated.Results: A high correlation in AFs was observed, as measured by the new and conventional manual methods. AF and RF scatterplots divided the SEMSs into three subgroups according to structure: hook-and-cross-type (low AF and RF), cross-type (high AF and low RF), and laser-cut-type (intermediate AF and high RF). The hook-and-cross-type had the largest axial force zero border (>20°), followed by the laser-cut and cross types.Conclusions: MPs were related to stent structure. Hook-and-cross-type SEMSs had a low AF and high axial force zero border and were considered safest because they caused minimal stress on the biliary wall. However, the increase in RF must be overcome.Keywords
This publication has 20 references indexed in Scilit:
- Understanding the Mechanical forces of Self-Expandable Metal Stents in the Biliary DuctsCurrent Gastroenterology Reports, 2016
- A fully covered self-expandable metal stent anchored by a 10-Fr double pigtail plastic stent: an effective anti-migration techniqueAnnals of Gastroenterology, 2016
- Self-expandable metal stents versus plastic stents for malignant biliary obstruction: a meta-analysisGastrointestinal Endoscopy, 2015
- Risk factors for covered metallic stent migration in patients with distal malignant biliary obstruction due to pancreatic cancerJournal of Gastroenterology and Hepatology, 2014
- Metallic stent with high axial force as a risk factor for cholecystitis in distal malignant biliary obstructionJournal of Gastroenterology and Hepatology, 2014
- Comparison of partially covered nitinol stents with partially covered stainless stents as a historical control in a multicenter study of distal malignant biliary obstruction: the WATCH studyGastrointestinal Endoscopy, 2012
- Risk factors for pancreatitis following transpapillary self-expandable metal stent placementSurgical Endoscopy, 2011
- Comparison of axial force and cell width of self‐expandable metallic stents: which type of stent is better suited for hilar biliary strictures?Journal of Hepato-Biliary-Pancreatic Sciences, 2011
- Measurement of radial and axial forces of biliary self-expandable metallic stentsGastrointestinal Endoscopy, 2009
- [Biliary endoprosthesis using Gore-Tex covered expandable metallic stents: preliminary clinical evaluation].1994