    Effect of Pulsed Field Timing Sequence and Electrode Cooling on Ablation Safety and Efficacy Surrogates: A Preclinical Study

    by Luigi Di Biase, Jonathan Hsu, Rahul Bhardwaj, David Newton, William Maddox, Jorge E. Romero, Tara Gomez, Eric Byun, Qi Chen, Christopher Beeckler, et al.

    BACKGROUND: The variable loop circular catheter has been shown to be a safe, effective pulsed field ablation catheter for atrial fibrillation treatment. After the commercial launch of the initial waveform (Seq1), a new waveform (Seq2 30 mL/min ) was developed that reduces electrode surface temperature and ablation delivery time (20–3.8 seconds). This study compares the safety and efficacy of Seq1 versus Seq2 across 2 irrigation rates and 2 dosing conditions. METHODS: Twenty-four Yorkshire pigs were treated with pulsed field ablation using the variable loop circular catheter. Subjects were randomized into 3 pulse recipe/irrigation arms (Seq1 4 mL/min , Seq1 30 mL/min , and Seq 2 30 mL/min ). Each group was split into a 1× nominal and a 2×-dose group, resulting in 6 total study groups. Animals were monitored for 28±2 days postablation for follow-up remapping procedures before euthanization. Gross necropsy and histopathology were performed on day 28±2. RESULTS: In all groups, 100% acute and chronic pulmonary vein isolation effectiveness was observed, as indicated by transmural, circumferential, pulmonary vein lesions with no evidence of collateral injury. Transmurality in the left atrial appendage, LA roof, and posterior wall was 75% to 100% across all groups and absent at the mitral isthmus. No catheter thrombus/char was observed at Seq2 30 mL/min or Seq1 30 mL/min . With Seq1 4 mL/min , thrombus/char occurred in 25% (1×) and 100% (2×). Hemolysis and myocardial injury marker changes were similar between all groups, with a notable increase in the 2× groups. On echocardiography, Seq2 30 mL/min microbubbles were rated as few or none, while Seq1 4 mL/min and Seq1 30 mL/min were mostly rated moderate or shower. CONCLUSIONS: Consistently durable pulmonary vein isolation was observed across all groups, including Seq1 4 mL/min , Seq2 30 mL/min , and Seq1 30 mL/min across both dosing conditions. Seq2 30 mL/min was associated with a reduced ablation delivery time (3.8 seconds), zero catheter thrombus formation, and few microbubbles. Higher irrigation flow in Seq1 30 mL/min also mitigated catheter thrombus formation but did not reduce microbubble generation.

  • Open Access

    Pre-clinical impact of higher irrigation flow rates during pulsed field ablation with a variable loop circular catheter

    by Fengwei Zou, Xiao-Dong Zhang, Tara Gomez, Eric Byun, Qi Chen, Jacopo Marazzato, Marco Schiavone, Sanghamitra Mohanty, Vincenzo Mirco La Fazia, Jishu Motta, et al.


  • Open Access

    Application repetition and electrode–tissue contact result in deeper lesions using a pulsed-field ablation circular variable loop catheter

    by Luigi Di Biase, Jacopo Marazzato, Tara Gomez, Eric Byun, Fengwei Zou, Vito Grupposo, Sanghamitra Mohanty, Vincenzo Mirco La Fazia, Giuseppe Ammirati, Aung Lin, et al.

    Pulsed-field ablation (PFA) is a novel, myocardial-selective, non-thermal ablation modality used to target cardiac arrhythmias. Although prompt electrogram (EGM) signal disappearance is observed immediately after PFA application in the pulmonary veins, whether this finding results in adequate transmural lesions is unknown. The aim of this study is to check whether application repetition and catheter–tissue contact impact lesion formation during PFA. A circular loop PFA catheter was used to deliver repeated energy applications with various levels of contact force. A benchtop vegetal potato model and a beating heart ventricular myocardial model were utilized to evaluate the impact of application repetition, contact force, and catheter repositioning on contiguity and lesion depth. Lesion development occurred over 18 h in the vegetal model and over 6 h in the porcine model. Lesion formation was found to be dependent on application repetition and contact. In porcine ventricles, single and multiple stacked applications led to a lesion depth of 3.5 ± 0.7 and 4.4 ± 1.3 mm, respectively (P = 0.002). Furthermore, the greater the catheter–tissue contact, the more contiguous and deeper the lesions in the vegetal model (1.0 ± 0.9 mm with no contact vs. 5.4 ± 1.4 mm with 30 g of force; P = 0.0001). Pulsed-field ablation delivered via a circular catheter showed that both repetition and catheter contact led independently to deeper lesion formation. These findings indicate that endpoints for effective PFA are related more to PFA biophysics than to mere EGM attenuation.