Application repetition and electrode–tissue contact result in deeper lesions using a pulsed-field ablation circular variable loop catheter
Authors:
- Luigi Di BiaseLBLuigi Di Biase
- Jacopo MarazzatoJMJacopo Marazzato
- Tara GomezTGTara Gomez
- Eric ByunEBEric Byun
- Fengwei ZouFZFengwei Zou
- Vito GrupposoVGVito Grupposo
- Sanghamitra MohantySMSanghamitra Mohanty
- Vincenzo Mirco La FaziaVFVincenzo Mirco La Fazia
- Giuseppe AmmiratiGAGiuseppe Ammirati
- Aung LinALAung Lin
- Domenico Della RoccaDRDomenico Della Rocca
Open Access
- 16 August 2024
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 26 (9)
Abstract
Pulsed-field ablation (PFA) is a novel, myocardial-selective, non-thermal ablation modality used to target cardiac arrhythmias. Although prompt electrogram (EGM) signal disappearance is observed immediately after PFA application in the pulmonary veins, whether this finding results in adequate transmural lesions is unknown. The aim of this study is to check whether application repetition and catheter–tissue contact impact lesion formation during PFA. A circular loop PFA catheter was used to deliver repeated energy applications with various levels of contact force. A benchtop vegetal potato model and a beating heart ventricular myocardial model were utilized to evaluate the impact of application repetition, contact force, and catheter repositioning on contiguity and lesion depth. Lesion development occurred over 18 h in the vegetal model and over 6 h in the porcine model. Lesion formation was found to be dependent on application repetition and contact. In porcine ventricles, single and multiple stacked applications led to a lesion depth of 3.5 ± 0.7 and 4.4 ± 1.3 mm, respectively (P = 0.002). Furthermore, the greater the catheter–tissue contact, the more contiguous and deeper the lesions in the vegetal model (1.0 ± 0.9 mm with no contact vs. 5.4 ± 1.4 mm with 30 g of force; P = 0.0001). Pulsed-field ablation delivered via a circular catheter showed that both repetition and catheter contact led independently to deeper lesion formation. These findings indicate that endpoints for effective PFA are related more to PFA biophysics than to mere EGM attenuation.Keywords
Funding Information
- Biosense Webster
This publication has 14 references indexed in Scilit:
- State-of-the-art pulsed field ablation for cardiac arrhythmias: ongoing evolution and future perspectiveEP Europace, 2024
- Initial clinical experience with the balloon-in-basket pulsed field ablation system: acute results of the VOLT CE mark feasibility studyEP Europace, 2024
- 2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society expert consensus statement on catheter and surgical ablation of atrial fibrillationEP Europace, 2024
- First-in-human clinical series of a novel conformable large-lattice pulsed field ablation catheter for pulmonary vein isolationEP Europace, 2024
- Comprehensive dose–response study of pulsed field ablation using a circular catheter compared with radiofrequency ablation for pulmonary vein isolation: A preclinical studyHeart Rhythm O2, 2023
- Determinants of acute irreversible electroporation lesion characteristics after pulsed field ablation: the role of voltage, contact, and adipose interferenceEP Europace, 2023
- Progress in atrial fibrillation ablation during 25 years of Europace journalEP Europace, 2023
- In vivopulsed-field ablation in healthy vs. chronically infarcted ventricular myocardium: biophysical and histologic characterizationEP Europace, 2023
- In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applicationsJournal of Cardiovascular Electrophysiology, 2022
- Circular Multielectrode Pulsed Field Ablation Catheter Lasso Pulsed Field AblationCirculation: Arrhythmia and Electrophysiology, 2021