Franz-Xaver Anzinger
Franz-Xaver Anzinger has a total of 11 co-authors. Here are the top co-authors by number of shared publications:
- Andreas Lindner
Department of Pediatric Surgery, Dr. von Hauner Children's Hospital, Ludwig-Maximilians-University of Munich, Lindwurmstr.4, 80337 Munich, Germany
2 shared publications - Jan Goedeke
Department of Pediatric Surgery, Dr. von Hauner Children’s Hospital, LMU University Hospital, Munich, Munich, Germany
2 shared publications M. Riboldi
Ludwig‐Maximilians‐University of Munich | LMU, Chair for Experimental Physics Munich Germany
2 shared publicationsO. J. Muensterer
Department of Pediatric Surgery Ludwig‐Maximilians‐University of Munich | LMU Munich Germany
2 shared publicationsP. Palaniappan
Ludwig‐Maximilians‐University of Munich | LMU, Chair for Experimental Physics Munich Germany
2 shared publicationsT. J. Hashagen
Ludwig‐Maximilians‐University of Munich | LMU, Chair for Experimental Physics Munich Germany
2 shared publications
Latest Publications
- Open Access
- 11 June 2026
- Published by Wolters Kluwer Health in Annals of Medicine & Surgery
- Vol. 88 (7) , 4663-4668
Double, double toil and trouble: a case of combined esophageal and gastrointestinal duplication cysts in infancy and a review of the literature
Introduction and importance: Gastrointestinal duplication cysts are rare congenital anomalies, usually solitary and most often involving the ileum and esophagus. Multiple non-contiguous duplications are exceptionally uncommon and may mimic more frequent neonatal thoracic or abdominal conditions. Case presentation: We report a term female neonate with a prenatal suspicion of congenital pulmonary airway malformation. Postnatally, she developed respiratory distress due to a large thoracic cyst, which was resected thoracoscopically and confirmed to be an esophageal duplication cyst. Weeks later, recurrent melena and anemia prompted re-evaluation. A Tc-99m pertechnetate scan suggested Meckel’s diverticulum, but laparoscopy revealed a 35 cm tubular ileal duplication with intestinal non-rotation. Resection and Ladd’s procedure were performed, with an uneventful recovery. Histopathology demonstrated gastric metaplasia and pancreatic heterotopia. Clinical discussion: This case highlights the diagnostic difficulty of non-contiguous foregut and midgut duplications presenting metachronously with respiratory distress and gastrointestinal bleeding. Ultrasound remains the preferred first-line modality, with CT, MRI, and Tc-99m pertechnetate scintigraphy used selectively. Complete excision is generally recommended, while suspected residual ectopic gastric mucosa may justify temporary acid suppression and close surveillance when the child is stable and reoperation is not immediately indicated. Conclusions: Gastrointestinal duplications should be considered in neonates with unexplained respiratory or gastrointestinal symptoms, particularly when associated anomalies are present. Multimodal imaging, timely surgery, and structured follow-up are important to prevent recurrent bleeding, obstruction, or other complications.
- 11 June 2026
- Published by Wolters Kluwer Health in Annals of Medicine & Surgery
- Vol. 88 (7) , 4663-4668
Open AccessDouble, double toil and trouble: a case of combined esophageal and gastrointestinal duplication cysts in infancy and a review of the literature
Introduction and importance: Gastrointestinal duplication cysts are rare congenital anomalies, usually solitary and most often involving the ileum and esophagus. Multiple non-contiguous duplications are exceptionally uncommon and may mimic more frequent neonatal thoracic or abdominal conditions. Case presentation: We report a term female neonate with a prenatal suspicion of congenital pulmonary airway malformation. Postnatally, she developed respiratory distress due to a large thoracic cyst, which was resected thoracoscopically and confirmed to be an esophageal duplication cyst. Weeks later, recurrent melena and anemia prompted re-evaluation. A Tc-99m pertechnetate scan suggested Meckel’s diverticulum, but laparoscopy revealed a 35 cm tubular ileal duplication with intestinal non-rotation. Resection and Ladd’s procedure were performed, with an uneventful recovery. Histopathology demonstrated gastric metaplasia and pancreatic heterotopia. Clinical discussion: This case highlights the diagnostic difficulty of non-contiguous foregut and midgut duplications presenting metachronously with respiratory distress and gastrointestinal bleeding. Ultrasound remains the preferred first-line modality, with CT, MRI, and Tc-99m pertechnetate scintigraphy used selectively. Complete excision is generally recommended, while suspected residual ectopic gastric mucosa may justify temporary acid suppression and close surveillance when the child is stable and reoperation is not immediately indicated. Conclusions: Gastrointestinal duplications should be considered in neonates with unexplained respiratory or gastrointestinal symptoms, particularly when associated anomalies are present. Multimodal imaging, timely surgery, and structured follow-up are important to prevent recurrent bleeding, obstruction, or other complications.
- Open Access
- 11 April 2025
- Published by Wiley in Pediatric Pulmonology
- Vol. 60 (4) , e71090
Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model
Background: In recent years, digital chest tube drainage systems have been introduced. Limited studies address their benefits and risks in pediatric patients, particularly neonates. This study compares a three‐chamber chest drainage system with a digital system using a high‐fidelity 3D‐printed model.Methods: We conducted direct measurements and 3D‐printed model tests with both systems at different suction pressures (−1 to −20 cmH2O) to assess the actual pressures. The effects of siphon and automatic flushes in the digital system were also studied.Results: At −20 and −10 cmH2O, significant differences were found between the digital and analog systems in direct and model measurements. The analog system became unreliable below −10 cmH2O. For the digital system, most measurements remained within the set pressures, with outliers up to −30 cmH2O due to regular flushing.Conclusion: This experimental study evaluates the suitability of digital drainage systems for the pediatric and neonatal populations. Our model demonstrated reliable simulation of thoracic conditions, making it a useful tool for pre‐clinical testing where patient testing may be limited. Both systems yielded satisfactory results at −20 and −10 cmH2O, but the digital system showed greater flexibility, maintaining pressures as low as −5 cmH2O. The analog system was consistent but less adaptable, which may limit its use in dynamic situations. The digital system's ability to simulate more flexible scenarios offers potential clinical advantages, though further investigation is needed to assess its impact on neonatal safety. The increase in suction during flushing may pose a risk for neonatal patients.
- 11 April 2025
- Published by Wiley in Pediatric Pulmonology
- Vol. 60 (4) , e71090
Open AccessComparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model
Background: In recent years, digital chest tube drainage systems have been introduced. Limited studies address their benefits and risks in pediatric patients, particularly neonates. This study compares a three‐chamber chest drainage system with a digital system using a high‐fidelity 3D‐printed model.Methods: We conducted direct measurements and 3D‐printed model tests with both systems at different suction pressures (−1 to −20 cmH2O) to assess the actual pressures. The effects of siphon and automatic flushes in the digital system were also studied.Results: At −20 and −10 cmH2O, significant differences were found between the digital and analog systems in direct and model measurements. The analog system became unreliable below −10 cmH2O. For the digital system, most measurements remained within the set pressures, with outliers up to −30 cmH2O due to regular flushing.Conclusion: This experimental study evaluates the suitability of digital drainage systems for the pediatric and neonatal populations. Our model demonstrated reliable simulation of thoracic conditions, making it a useful tool for pre‐clinical testing where patient testing may be limited. Both systems yielded satisfactory results at −20 and −10 cmH2O, but the digital system showed greater flexibility, maintaining pressures as low as −5 cmH2O. The analog system was consistent but less adaptable, which may limit its use in dynamic situations. The digital system's ability to simulate more flexible scenarios offers potential clinical advantages, though further investigation is needed to assess its impact on neonatal safety. The increase in suction during flushing may pose a risk for neonatal patients.
- Open Access
- 19 November 2024
- Published by Wiley
Comparative Analysis of Analog and Digital Thoracic Drainage Systems Using a High-Fidelity 3D-Printed Neonatal Chest Phantom
Background: In recent years, digital thoracic drainage systems have been introduced. Limited studies address the benefits and risks in pediatric patients. We compared analog and digital systems using a high-fidelity 3D-printed phantom. Methods: Direct measurements as well as measurements on a 3D-printed phantom were conducted employing a digital and analog system for different suction pressures (-1 to -20 cmH 2 O). The impact of a siphon and automatic flushes of the digital system were studied. Results: At pressure settings of -20 and -10 cmH 2 O, direct and phantom measurements yielded significantly different results for the digital compared to the analog system. Set pressures below -10 cmH 2 O in the analog drainage system were unreliable. For the digital system, the majority of measured points fell within the chosen settings. There were statistical outliers up to -30 cmH 2 O attributed to the regular flushing performed by the system. Conclusion: This is the first experimental study objectifying the suitability of digital drainage systems for the pediatric and neonatal population. Our phantom produced accurate results where tests on patients are not possible. At suction pressures of -20 and -10 cmH 2 O the digital and the analog drainage system yield satisfactory results. The digital system was able to maintain pressures as low as -5 cmH 2 O. The analog drainage system is less flexible but remains a reliable tool for experienced users. The digital drainage system allows more freedom and can increase patient safety by neutralizing obstructions or a siphon. The increase in suction during flushing may pose a risk for neonatal patients.
- 19 November 2024
- Published by Wiley
Open AccessComparative Analysis of Analog and Digital Thoracic Drainage Systems Using a High-Fidelity 3D-Printed Neonatal Chest Phantom
Background: In recent years, digital thoracic drainage systems have been introduced. Limited studies address the benefits and risks in pediatric patients. We compared analog and digital systems using a high-fidelity 3D-printed phantom. Methods: Direct measurements as well as measurements on a 3D-printed phantom were conducted employing a digital and analog system for different suction pressures (-1 to -20 cmH 2 O). The impact of a siphon and automatic flushes of the digital system were studied. Results: At pressure settings of -20 and -10 cmH 2 O, direct and phantom measurements yielded significantly different results for the digital compared to the analog system. Set pressures below -10 cmH 2 O in the analog drainage system were unreliable. For the digital system, the majority of measured points fell within the chosen settings. There were statistical outliers up to -30 cmH 2 O attributed to the regular flushing performed by the system. Conclusion: This is the first experimental study objectifying the suitability of digital drainage systems for the pediatric and neonatal population. Our phantom produced accurate results where tests on patients are not possible. At suction pressures of -20 and -10 cmH 2 O the digital and the analog drainage system yield satisfactory results. The digital system was able to maintain pressures as low as -5 cmH 2 O. The analog drainage system is less flexible but remains a reliable tool for experienced users. The digital drainage system allows more freedom and can increase patient safety by neutralizing obstructions or a siphon. The increase in suction during flushing may pose a risk for neonatal patients.