WZWeiting Mm Zhang

Weiting Mm Zhang

*This is an algorithmically generated author record based on publicly available information.

Current Organizations / Institutes

The organizations/institutes from the scholar's latest article


Weiting Mm Zhang has a total of 9 co-authors. Here are the top co-authors by number of shared publications:

  • Dian Mm Yuan

    The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou

    1 shared publication
  • Jianbo Gao

    The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou

    1 shared publication
  • Jie Liu

    Department of Radiology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China

    1 shared publication
  • Ke Mm Qi

    The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou

    1 shared publication
  • Linfeng Li

    Siemens Healthineers, Shanghai, China

    1 shared publication
  • Mengyuan Mm Zhang

    The Department of Radiology, The First Affiliated Hospital of Zhengzhou, University, Zhengzhou

    1 shared publication

Latest Publications

  • Feasibility Analysis of Individualized Low Flow Rate Abdominal Contrast-Enhanced Computed Tomography in Chemotherapy Patients: Dual-Source Computed Tomography With Low Tube Voltage

    by Yicun Mm Zhang, Dian Mm Yuan, Ke Mm Qi, Mengyuan Mm Zhang, Weiting Mm Zhang, Nannan Mm Wei, Linfeng Li, Peijie Lv, Jianbo Gao, Jie Mm Liu

    Purpose: The aim of the study is to investigate the feasibility of using dual-source computed tomography (CT) combined with low flow rate and low tube voltage for postchemotherapy image assessment in cancer patients. Methods: Ninety patients undergoing contrast-enhanced CT scans of the upper abdomen were prospectively enrolled and randomly assigned to groups A, B, and C (n = 30 each). In group A, patients underwent scans at 120 kVp with 448 mgI/kg. Patients in group B underwent scans at 100 kVp with 336 mgI/kg. Patient in group C underwent scans at 70 kVp with of 224 mgI/kg. Quantitative measurements including the CT number, standard deviation of CT number, signal-to-noise ratio, contrast-to-noise ratio, subjective reader scores, and the volume and flow rate of contrast agent were evaluated for each group. Results: There was no statistically significant difference in the subjective image scores within the three groups except for the kidney (all P > 0.05). Group C showed significantly higher CT values, lower noise levels, and higher signal-to-noise ratio and contrast-to-noise ratio values in the majority of the regions of interest compared to the other groups (P < 0.05). In group C, the contrast agent dose was decreased by 46% compared to group A (79.48 ± 12.24 vs 42.7 ± 8.6, P < 0.01), and the contrast agent injection rate was reduced by 22% (2.7 ± 0.41 vs 2.1 ± 0.4, P < 0.01). Conclusions: The use of 70 kVp tube voltage combined with low iodine flow rates prove to be a more effective approach in solving the challenge of compromised blood vessels in postchemotherapy tumor patients, without reducing image quality and diagnostic confidence.