Towards optimal imaging with PET: anin silicofeasibility study
Authors:
- A L McNamaraAMA L McNamara
- M ToghyaniMTM Toghyani
- J E GillamJGJ E Gillam
- K WuKWK Wu
- Z KuncicZKZ Kuncic
- 21 November 2014
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (24) , 7587-7600
Abstract
The efficacy of Positron Emission Tomography (PET) imaging relies fundamentally on the ability of the system to accurately identify true coincidence events. With existing systems, this is currently accomplished with an energy acceptance criterion followed by correction techniques to remove suspected false coincidence events. These corrections generally result in signal and contrast loss and thus limit the PET system's ability to achieve optimum image quality. A key property of annihilation radiation is that the photons are polarised with respect to each other. This polarisation correlation offers a potentially powerful discriminator, independent of energy, to accurately identify true events. In this proof of concept study, we investigate how photon polarisation information can be exploited in PET imaging by developing a method to discriminate true coincidences using the polarisation correlation of annihilation pairs. We implement this method using a Geant4 PET simulation of a GE Advance/Discovery LS system and demonstrate the potential advantages of the polarisation coincidence selection method over a standard energy criterion method. Current PET ring detectors are not capable of exploiting the polarisation correlation of the photon pairs. Compton PET systems, however are promising candidates for this application. We demonstrate the feasibility of a two-component Compton camera system in identifying true coincidences with Monte Carlo simulations. Our study demonstrates the potential of improving signal gain using polarisation, particularly for high photon emission rates. We also demonstrate the ability of the Compton camera at exploiting this polarisation correlation in PET.Keywords
This publication has 35 references indexed in Scilit:
- Silicon as an unconventional detector in positron emission tomographyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- FMT-XCT: in vivo animal studies with hybrid fluorescence molecular tomography–X-ray computed tomographyNature Methods, 2012
- Performance evaluation of a pixellated Ge Compton cameraPhysics in Medicine & Biology, 2011
- A Pixelated Emission Detector for RadiOisotopes (PEDRO)Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2009
- Optimizing Compton camera geometriesPhysics in Medicine & Biology, 2004
- In vivomolecular and genomic imaging: new challenges for imaging physicsPhysics in Medicine & Biology, 2004
- New Monte Carlo method for Compton and Rayleigh scattering by polarized gamma raysNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Polarization constraints on gamma-ray event circles in compton scatter instrumentsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Correlation between the States of Polarization of the Two Quanta of Annihilation RadiationPhysical Review B, 1948