Vascular Access: Comparison of US Guidance with the Sonic Flashlight and Conventional US in Phantoms
Authors:
- Wilson M. ChangWCWilson M. Chang
- Nikhil B. AmesurNANikhil B. Amesur
- Roberta L. KlatzkyRKRoberta L. Klatzky
- Albert B. ZajkoAZAlbert B. Zajko
- George D. StettenGSGeorge D. Stetten
- 1 December 2006
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 241 (3) , 771-779
Abstract
To prospectively evaluate whether ultrasonography (US)-guided vascular access can be learned and performed faster with the sonic flashlight than with conventional US and to demonstrate sonic flashlight-guided vascular access in a cadaver. Institutional review board approval and oral and written informed consent were obtained. The sonic flashlight replaces the standard US monitor with a real-time US image that appears to float beneath the skin and is displayed where it is scanned. In studies 1 and 2, participants performed sonic flashlight-guided needle insertion tasks in vascular phantoms. In study 1, 16 participants (nine women, seven men) with no US experience performed 60 simulated vascular access trials with sonic flashlight or conventional US guidance. With analysis of variance (ANOVA) and power-curve fitting, improvement with practice rate and mean differences between techniques and tasks were examined. In study 2, 14 female nurses (mean age, 50.1 years) proficient with conventional US performed simulated vascular access trials on three tasks with the sonic flashlight and conventional US. With random assignment, half the participants used the sonic flashlight first and half used conventional US first. Mean performance with each technique and that with each task were compared by using ANOVA. In study 3, feasibility of sonic flashlight guidance for access to internal jugular and basilic veins was demonstrated in a cadaver. For study 1, learning rates (ie, decrease in access time over trials) did not differ for vascular access with sonic flashlight and conventional US. Overall, participants achieved faster vascular access times with sonic flashlight guidance (P < .007). In study 2, participants performed procedures faster overall with the sonic flashlight (P < .02) and found the sonic flashlight easier to use. In study 3, sonic flashlight-guided vascular access was gained in the cadaver. Learning and performance of vascular access were significantly faster with the sonic flashlight than with conventional US, and vascular access could be gained in a cadaver; the sonic flashlight is ready for clinical trials.Keywords
This publication has 8 references indexed in Scilit:
- Robotic Long-distance Telementoring in NeurosurgeryNeurosurgery, 2005
- Guidance of Retrobulbar Injection With Real‐time Tomographic ReflectionJournal of Ultrasound in Medicine, 2002
- Augmented reality guidance for needle biopsies: An initial randomized, controlled trial in phantoms††A preliminary version of this paper was presented at the Medical Image Computing and Computer-Assisted Intervention (MICCAI) 2001 conference in Utrecht. The Netherlands (Rosenthal et al., 2001)Medical Image Analysis, 2002
- Ultrasound Guidance versus the Landmark Technique for the Placement of Central Venous Catheters in the Emergency DepartmentAcademic Emergency Medicine, 2002
- Use of ultrasound to place central linesJournal of Critical Care, 2002
- Overlaying ultrasonographic images on direct vision.Journal of Ultrasound in Medicine, 2001
- Sonographically guided placement of peripherally inserted central venous catheters: review of 355 procedures.American Journal of Roentgenology, 1998
- Real-time Ultrasonic Guidance for Percutaneous Puncture of the Internal Jugular VeinAnesthesiology, 1986