Abstract
Depth-of-interaction (DOI) capability is important for achieving high spatial resolution and sensitivity in dedicated organ and small animal positron emission tomography (PET) scanners. The dual-ended readout is one of the common methods that can achieve good DOI resolution. The aim of this study is to evaluate a dual-ended readout detector based on silicon photomultiplier (SiPM) and TOFPET2 application-specific integrated circuit (ASIC). The detector is based on 4 4 lutetium–yttrium oxyorthosilicate (LYSO) units, each unit contained 6 6 LYSO crystals, and the crystal size was 1 1 20 mm 3 . The four lateral surfaces of LYSO crystals were mechanically ground to W14 (surface roughness 10–14 m), and the two ended surfaces were polished (surface roughness 2 and 1.53 (at 420 nm), respectively. To characterize the detector performance at different depths, another 1 25.8 20 mm 3 single LYSO slab was used. Data were acquired at 10 depths (1, 3, …, 19 mm), and each depth had a 10 min acquisition time and about 40 thousand coincidence events. During the experiment, the SiPM temperature was controlled as 27.6 0.4 °C. The results showed that the 1.2 mm light guide offered the best crystal resolvability. The energy, coincidence time, and DOI resolution full-width at half-maximum of the detector were characterized as 15.66% 0.66%, 602.98 10.58 ps, and 2.33 0.07 mm, respectively. The good DOI resolution indicates the potential of utilizing the detector for high-resolution PET applications.
Funding Information
  • National Institute of Biomedical Imaging and Bioengineering (R01EB028091)

This publication has 29 references indexed in Scilit: