Characterization of a high-resolution hybrid DOI detector for a dedicated breast PET/CT scanner
- 11 May 2012
- journal article
- conference paper
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (11) , 3435-3449
Abstract
The aim of this study is to design and test a new high-resolution hybrid depth of interaction (DOI) detector for a dedicated breast PET/CT scanner. Two detectors have been designed and built. The completed detectors are based on a 14 × 14 array of 1.5 × 1.5 × 20 mm(3) unpolished lutetium orthosilicate scintillation crystals, with each element coated in a 50 μm layer of reflective material. The detector is read out from both ends using a position-sensitive photomultiplier tube (PSPMT) and a large active area (20 × 20 mm(2)) avalanche photodiode (APD) to enable acquisition of DOI information. Nuclear instrumentation modules were used to characterize the detectors' performances in terms of timing, intrinsic spatial resolution (ISR) and energy resolution, as well as DOI resolution with a dual-ended readout configuration. Measurements with the APD were performed at a temperature of 10 °C. All crystals were identified at all depths, even though the signal amplitude from the PSPMT decreases with depth away from it. We measured a timing resolution of 2.4 ns, and an average energy resolution of 19%. The mean ISR was measured to be 1.2 mm for crystals in the central row of the array for detectors in the face-to-face position. Two off-center positions were measured corresponding to 26° and 51° oblique photon incidence, and the mean ISR at these positions was 1.5 and 1.7 mm, respectively. The average DOI resolution across all crystals and depths was measured to be 2.9 mm (including the beam width of 0.6 mm). This detector design shows good promise as a high-resolution detector for a dedicated breast PET/CT scanner.Keywords
This publication has 33 references indexed in Scilit:
- Initial clinical test of a breast-PET scannerJournal of Medical Imaging and Radiation Oncology, 2011
- Tapered LSO arrays for small animal PETPhysics in Medicine & Biology, 2010
- Clinical Imaging Characteristics of the Positron Emission Mammography Camera: PEM Flex Solo IIJournal of Nuclear Medicine, 2009
- Initial Characterization of a Dedicated Breast PET/CT Scanner During Human ImagingJournal of Nuclear Medicine, 2009
- PET characteristics of a dedicated breast PET/CT scanner prototypePhysics in Medicine & Biology, 2009
- Crystal identification in positron emission tomography using nonrigid registration to a Fourier-based templatePhysics in Medicine & Biology, 2008
- A Prototype PET Scanner with DOI-Encoding DetectorsJournal of Nuclear Medicine, 2008
- Virtual-Pinhole PETJournal of Nuclear Medicine, 2008
- Analysis of factors affecting positron emission mammography (PEM) image formationIEEE Transactions on Nuclear Science, 2003
- A room temperature LSO/PIN photodiode PET detector module that measures depth of interactionIEEE Transactions on Nuclear Science, 1995