Comparative Clinical Study of Surgical Performance of Quatera 700 versus Centurion and Signature Pro Phacoemulsification Systems
Open Access
- 1 September 2024
- journal article
- research article
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. ume 18, 2685-2695
Abstract
Purpose: To compare the intra-operative surgical performance of Quatera 700 with Centurion and Signature Pro phacoemulsification systems. Setting: Nethradhama Superspeciality Eye Hospital, Bangalore, India. Study Design: Prospective, interventional, 3-arm, randomized comparison study. Methods: A total of 180 non-consecutive eyes satisfying the eligibility criteria were recruited. N = 60 eyes were operated using each of the above systems, of which n = 30 eyes had LOCS III grade 1– 2, and n = 30 eyes had grade 3– 4 nuclear density. Phacoemulsification with foldable IOL implantation was performed by a single experienced surgeon. Chamber stability scores (1– 4, 4 being the worst), fluid turnover per case, surge incidence and intra-operative complications were assessed. Results: Intra-operatively, the mean chamber stability score was significantly better for the Quatera 700 in both the soft and dense cataract groups compared to the other two machines (p < 0.001). Total Phaco Time (TPT) with Quatera 700 was significantly lower compared to Signature Pro, while it was comparable with Centurion for both the soft and dense cataract groups. The mean fluid turnover was significantly lower for the Quatera 700 in the dense cataract group (p = 0.04) and so was the combined phaco and I/A time (p = 0.01) when compared with Signature Pro. One eye operated with Signature Pro and 2 eyes operated with Centurion had PCR due to surge. The mean corneal clarity scores on POD-1 and specular counts at 2 weeks and 6 months were comparable. Conclusion: The new Quatera 700 system resulted in lesser TPT, fluid turnover and phaco + I/A time than Signature Pro, while its performance was comparable to Centurion for these parameters. Moreover, the surgeon graded intra-operative chamber stability scoring was highest, when compared to the other two systems. Ctri Registration Number: CTRI/2022/01/039296.Keywords
This publication has 28 references indexed in Scilit:
- Experimental study comparing 2 different phacoemulsification systems with intraocular pressure control during steady-state flow and occlusion break surge eventsJournal of Cataract & Refractive Surgery, 2023
- Aqueous volume loss associated with occlusion break surge in phacoemulsifiers from 4 different manufacturersJournal of Cataract & Refractive Surgery, 2018
- Comparative clinical study of Whitestar Signature phacoemulsification system with standard and Ellips FX handpiecesInternational Ophthalmology, 2017
- Laser in situ keratomileusis for astigmatism ≤ 0.75 Diopter combined with low myopia: a retrospective data analysisBMC Ophthalmology, 2014
- Stroke dynamics and frequency of 3 phacoemulsification machinesJournal of Cataract & Refractive Surgery, 2012
- Comparison of endothelial changes and power settings between torsional and longitudinal phacoemulsificationJournal of Cataract & Refractive Surgery, 2010
- Standardization of Uveitis Nomenclature for Reporting Clinical Data. Results of the First International WorkshopAmerican Journal of Ophthalmology, 2005
- Flared phacoemulsification tips to decrease ultrasound time and energy in cataract surgeryJournal of Cataract & Refractive Surgery, 2001
- Ultrasonic energy in phacoemulsification: Mechanical cutting and cavitationJournal of Cataract & Refractive Surgery, 1994
- The Lens Opacities Classification System IIIArchives of Ophthalmology (1950), 1993