Performance evaluation of platelet counting of Abbott Alinity hq and Sysmex XN‐9000 automated hematology analyzer compared with international reference method
- 20 November 2020
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 43 (3) , 387-394
Abstract
Introduction: Accurate platelet counting is essential for risk assessment of bleeding and thrombosis. Abbott Alinity hq hematology analyzer was recently introduced, and its performance in platelet counting has yet to be evaluated comprehensively. In this study, we evaluated the performance of the optical platelet counting of Abbott Alinity hq (Alinity‐PLT) and the impedance and fluorescent platelet counting of Sysmex XN‐9000 (XN‐PLT‐I and XN‐PLT‐F) compared with the international reference method.Methods: Blood samples were analyzed via Alinity hq and XN‐9000 with PLT‐F channel. Immuno‐platelet (ImmnoPLT) reference method was performed with CD41/CD61 antibodies using FACSLyricTM flow cytometer (BD). Precision was determined using 10 replicates in a single run, and the platelet counts of Alinity‐PLT, XN‐PLT‐I, XN‐PLT‐F, and ImmnoPLT were compared.Results: At a platelet count of 13 × 109/L, the CVs of Alinity‐PLT, XN‐PLT‐I, and XN‐PLT‐F were 4.2%, 6.7%, and 4.3%, respectively, and at a platelet count of 44 × 109/L, all showed a CV of less than 3%. For the total 210 samples, all three methods showed a very strong correlation with ImmunoPLT (r > 0.99). For platelet levels below 20 × 109/L, XN‐PLT‐F showed the strongest correlation with ImmunoPLT (r = 0.975), and for platelet levels of 20‐100 × 109/L, Alinity‐PLT and XN‐PLT‐I were comparable to ImmunoPLT. For platelet levels of 100‐450 × 109/L, XN‐PLT‐I was the most comparable to ImmunoPLT, and for platelet levels above 450 × 109/L, Alinity‐PLT was comparable to ImmunoPLT.Conclusions: All three methods were highly correlated with ImmunoPLT, and each method had different performance advantages according to the platelet levels.Keywords
Funding Information
- Abbott Korea
This publication has 13 references indexed in Scilit:
- Performance evaluation of new Abbott Alinity hq hematology analyzerInternational Journal of Laboratory Hematology, 2019
- Multicenter performance evaluation of the Abbott Alinity hq hematology analyzerClinical Chemistry and Laboratory Medicine (CCLM), 2019
- Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet countingPLOS ONE, 2019
- Performance evaluation of the prototype Abbott Alinity hq hematology analyzerInternational Journal of Laboratory Hematology, 2019
- Guidelines for the use of platelet transfusionsBritish Journal of Haematology, 2016
- Platelet Transfusion: A Clinical Practice Guideline From the AABBAnnals of Internal Medicine, 2015
- The Sysmex XN-2000 Hematology Autoanalyzer Provides a Highly Accurate Platelet Count than the Former Sysmex XE-2100 System Based on Comparison with the CD41/CD61 Immunoplatelet Reference Method of Flow CytometryAnnals of Laboratory Medicine, 2014
- New Fluorescent Method (PLT-F) on Sysmex XN2000 Hematology Analyzer Achieved Higher Accuracy in Low Platelet CountingAmerican Journal of Clinical Pathology, 2013
- Platelet Counting by the RBC/Platelet Ratio MethodAmerican Journal of Clinical Pathology, 2001
- An Interlaboratory Study of a Candidate Reference Method for Platelet CountingAmerican Journal of Clinical Pathology, 2001