Abstract
A rapid calibration system is under development for the Near Backscatter Imager (NBI) in use at the National Ignition Facility (NIF). NBI is an optical diagnostic that quantifies the stimulated Brillouin and Raman backscatter produced by NIF’s targets. Specifically, NBI measures the light that does not fall directly back into the laser aperture, which is measured by the Full Aperture Backscatter System (FABS). When working in tandem with FABS, NBI allows for the full characterization of backscattered light. This informs Hohlraum laser coupling, optical damage, and laser–plasma interaction models. NBI uses a large Spectralon plate covered by a protective glass layer and is mounted inside the target chamber where it is exposed to high energy backscatter, neutrons, and build-up debris left over from the exploded targets. This gradually alters the reflectivity of the plate, meaning that NBI needs to be calibrated regularly. Described here is NIF’s design for a system capable of rapid in situ calibration of NBI that is to be installed in FY25.Keywords
Funding Information
- Lawrence Livermore National Laboratory (DE-AC52-07NA27344)
This publication has 15 references indexed in Scilit:
- Observations and properties of the first laboratory fusion experiment to exceed a target gain of unityPhysical Review E, 2024
- Design of the first fusion experiment to achieve target energy gainPhysical Review E, 2024
- Optimization of Backscatter and Symmetry for Laser Fusion Experiments Using Multiple Tunable WavelengthsPhysical Review Applied, 2022
- The Scattered Light Time-history Diagnostic suite at the National Ignition FacilityReview of Scientific Instruments, 2021
- Time resolved detection of two-plasmon decay using three-halves harmonic emission on the National Ignition FacilityReview of Scientific Instruments, 2018
- The relationship between gas fill density and hohlraum drive performance at the National Ignition FacilityPhysics of Plasmas, 2017
- Three-wavelength scheme to optimize hohlraum coupling on the National Ignition FacilityPhysical Review E, 2011
- Backscatter measurements for NIF ignition targets (invited)Review of Scientific Instruments, 2010
- Gas-filled hohlraum experiments at the National Ignition FacilityPhysics of Plasmas, 2006
- Implementation of a near backscattering imaging system on the National Ignition FacilityReview of Scientific Instruments, 2004