Performance measurements of a depth-encoding TraPET detector module
Authors:
- 1 November 2011
- journal article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 6 (11) , C11032
Abstract
The new method to correct a parallax error and the loss of coincidence counts caused by the gap between modules was developed for a small animal PET scanner. We proposed the TraPET scanner composed of 6 dual-layer phoswich detector modules. Each detector module consists of a 5.0 mm-thick trapezoidal-shaped monolithic LSO crystal and a 23 x 23 array of GSO crystal. The layer of interaction is identified by the pulse shape discrimination method. One detector module was built and the algorithm for layer identification was optimized. The dual-layer crystals were optically coupled to a Hamamatsu H8500 position-sensitive PMT and a resistive charge divider was used to multiplex 64-channel anode outputs into 4-channel position signals. The 4 signals have been sampled continuously by 14-bit ADC at a sampling rate of 105 MHz and the pulse shape discrimination algorithm was achieved through FPGA programming. The detector module was irradiated with a Na-22 point source from the side of the crystals to obtain flood images of each layer and two layers were clearly identified, thus verifying the DOI capability. The TraPET detector proved to be a reliable design for correcting the parallax error and improving the sensitivity simultaneously in the small animal PET.Keywords
This publication has 7 references indexed in Scilit:
- New design of a quasi-monolithic detector module with DOI capability for small animal petNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2008
- Scintillator studies for the HPD-PET conceptNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007
- Evaluation of static physics performance of the jPET-D4 by Monte Carlo simulationsPhysics in Medicine & Biology, 2006
- Characterization of dual layer phoswich detector performance for small animal PET using Monte Carlo simulationPhysics in Medicine & Biology, 2004
- Correction methods for missing data in sinograms of the HRRT PET scanner2004
- Pulse shape discrimination of LSO and LuYAP scintillators for depth of interaction detection in PETIEEE Transactions on Nuclear Science, 2003
- Constrained Fourier space method for compensation of missing data in emission computed tomographyIEEE Transactions on Medical Imaging, 1988