Development and biomechanical analysis of an axially controlled compression spinal rod for lumbar spondylolysis
Open Access
- 7 June 2024
- journal article
- research article
- Published by Wolters Kluwer Health in Medicine
- Vol. 103 (23) , e38520
Abstract
Background: To elucidate the differences in mechanical performance between a novel axially controlled compression spinal rod (ACCSR) for lumbar spondylolysis (LS) and the common spinal rod (CSR). Methods: A total of 36 ACCSRs and 36 CSRs from the same batch were used in this study, each with a diameter of 6.0 mm. Biomechanical tests were carried out on spinal rods for the ACCSR group and on pedicle screw-rod internal fixation systems for the CSR group. The spinal rod tests were conducted following the guidelines outlined in the American Society for Testing and Materials (ASTM) F 2193, while the pedicle screw-rod internal fixation system tests adhered to ASTM F 1798-97 standards. Results: The stiffness of ACCSR and CSR was 1559.15 ± 50.15 and 3788.86 ± 156.45 N/mm (P < .001). ACCSR’s yield load was 1345.73 (1297.90–1359.97) N, whereas CSR’s was 4046.83 (3805.8–4072.53) N (P = .002). ACCSR’s load in the 2.5 millionth cycle of the fatigue four-point bending test was 320 N. The axial gripping capacity of ACCSR and CSR was 1632.53 ± 165.64 and 1273.62 ± 205.63 N (P = .004). ACCSR’s torsional gripping capacity was 3.45 (3.23–3.47) Nm, while CSR’s was 3.27 (3.07–3.59) Nm (P = .654). The stiffness of the pedicle screws of the ACCSR and CSR group was 783.83 (775.67–798.94) and 773.14 (758.70–783.62) N/mm (P = .085). The yield loads on the pedicle screws of the ACCSR and CSR group was 1345.73 (1297.90–1359.97) and 4046.83 (3805.8–4072.53) N (P = .099). Conclusion: Although ACCSR exhibited lower yield load, stiffness, and fatigue resistance compared to CSR, it demonstrated significantly higher axial gripping capacity and met the stress requirement of the human isthmus. Consequently, ACCSR presents a promising alternative to CSR for LS remediation.Keywords
This publication has 31 references indexed in Scilit:
- Mechanical Analysis of Notch-Free Pre-Bent Rods for Spinal Deformity SurgerySpine, 2020
- Spondylolysis 2019 updateCurrent Opinion in Pediatrics, 2019
- Experimental model for controlling shear using the Ilizarov frameClinical Biomechanics, 2015
- Biomechanical testing of a unique built-in expandable anterior spinal internal fixation systemBMC Musculoskeletal Disorders, 2014
- Measurement of the number of lumbar spinal movements in the sagittal plane in a 24-hour periodEuropean Spine Journal, 2014
- Double tension band osteosynthesis in transverse supracondylar distal humerus fractures and nonunionsInternational Orthopaedics, 2012
- Intraoperative Techniques to Reduce the Potential of Set-screw Loosening in Long Spinal ConstructsJournal of Spinal Disorders & Techniques, 2010
- Physical Characteristics of Polyaxial-Headed Pedicle Screws and Biomechanical Comparison of Load With Their FailureSpine, 2003
- Effects of Spinal Instrumentation on Fusion of the Lumbosacral SpineSpine, 1995