A technique for the reduction of RF‐induced heating of active implantable medical devices during MRI
- 10 August 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 87 (1) , 349-364
Abstract
Purpose The paper presents a novel method to reduce the RF-induced heating of active implantable medical devices during MRI. Methods With the addition of an energy decoying and dissipating structure, RF energy can be redirected toward the dissipating rings through the decoying conductor. Three lead groups (45 cm-50 cm) and 4 (50 cm-100 cm) were studied in 1.5 Tesla MR systems by simulation and measurement, respectively. In vivo modeling was performed using human models to estimate the RF-induced heating of an active implantable medical device for spinal cord treatment. Result In the simulation study, it was shown that the peak 1g-averaged specific absorption rate near the lead-tips can be reduced by 70% to 80% compared to those from the control leads. In the experimental measurements during a 2-min exposure test in a 1.5 Telsa MR system, the temperature rises dropped from the original 18.3℃, 25.8℃, 8.1℃, and 16.1℃ (control leads 1-4) to 5.4℃, 6.9℃, 1.6℃, and 3.3℃ (leads 1-4 with the energy decoying and dissipation structure). The in vivo calculation results show that the maximum induced temperature rise among all cases can be substantially reduced (up to 80%) when the energy decoying and dissipating structures were used. Conclusion Our studies confirm the effectiveness of the novel technique for a variety of scanning scenarios. The results also indicate that the decoying conductor length, number of rings, and ring area must be carefully chosen and validated.Keywords
Funding Information
- Division of Industrial Innovation and Partnerships (1922389)
This publication has 35 references indexed in Scilit:
- Reconfigurable MRI coil technology can substantially reduce RF heating of deep brain stimulation implants: First in-vitro study of RF heating reduction in bilateral DBS leads at 1.5 TPLOS ONE, 2019
- RF-induced heating in tissue near bilateral DBS implants during MRI at 1.5 T and 3T: The role of surgical lead managementNeuroImage, 2019
- Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implantsNeuroImage, 2017
- Feasibility of using linearly polarized rotating birdcage transmitters and close-fitting receive arrays in MRI to reduce SAR in the vicinity of deep brain simulation implantsMagnetic Resonance in Medicine, 2016
- Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance ImagingPLOS ONE, 2015
- Reduction of the radiofrequency heating of metallic devices using a dual‐drive birdcage coilMagnetic Resonance in Medicine, 2012
- Reduction of implant RF heating through modification of transmit coil electric fieldMagnetic Resonance in Medicine, 2010
- Measuring RF‐induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leadsMagnetic Resonance in Medicine, 2009
- Minimizing RF heating of conducting wires in MRIMagnetic Resonance in Medicine, 2007
- MRI and implanted medical devices: basic interactions with an emphasis on heatingIEEE Transactions on Device and Materials Reliability, 2005