Fidelity of dose delivery at high dose rate of volumetric modulated arc therapy in a truebeam linac with flattening filter free beams
Open Access
- 1 January 2012
- journal article
- research article
- Published by Medknow in Journal of Medical Physics
- Vol. 37 (4) , 193-9
Abstract
The purpose of this study is to assess fidelity of radiation delivery between high and low dose rates of the flattening filter free (FFF) modes of a new all-digital design medical linear accelerator (Varian TrueBeam TM ), particularly for plans optimized for volumetric modulated arc therapy (VMAT). Measurements were made for the two energies of flattening filter free photon beams with a Varian TrueBeam TM linac: 6 MV (6 XFFF) at 400 and 1400 MU/min, and 10 MV (10 XFFF) at 400 and 2400 MU/ min. Data acquisition and analysis was performed with both ionization chambers and diode detector system Delta 4 , for square radiation fields and for 8 VMAT treatment plans optimized for SBRT treatment of lung tumors. For the square fields, a percent dose difference between high and low dose rate of the order of 0.3-0.4% for both photon energies was seen with the ionization chambers, while the contribution to the difference from ion recombination was found to be negligible. For both the VMAT and square-field deliveries, the Delta 4 showed the same average percent dose difference between the two dose rates of ~0.8% and ~0.6% for 10 MV and 6 MV, respectively, with the lower dose rate values giving the greater measured dose compared to the high dose rate. Thus, the VMAT deliveries introduced negligible dose differences between high and low dose rate. Finally, reproducibility of dose measurements was good for both energies.Keywords
This publication has 34 references indexed in Scilit:
- Unflattened photon beams from the standard flattening filter free accelerators for radiotherapy: Advantages, limitations and challengesJournal of Medical Physics, 2011
- Experimental investigation of the response of an EPID to an unflattened photon beam from an Elekta Precise linear acceleratorMedical Physics, 2009
- The characterization of unflattened photon beams from a 6 MV linear acceleratorPhysics in Medicine & Biology, 2008
- Measurements of in-air output ratios for a linear accelerator with and without the flattening filterMedical Physics, 2006
- A flattening filter free photon treatment concept evaluation with Monte CarloMedical Physics, 2006
- Monte Carlo simulation of the photoneutron field in linac radiotherapy treatments with different collimation systemsPhysics in Medicine & Biology, 2004
- Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM codeMedical Physics, 2002
- Calculation of x‐ray spectra for radiosurgical beamsMedical Physics, 1995
- A Monte Carlo study of accelerator head scatterMedical Physics, 1994
- Buildup region and depth of dose maximum of megavoltage x-ray beamsMedical Physics, 1994