Corneal Power Measurement Obtained by Fourier-Domain Optical Coherence Tomography
Authors:
- Yanjun HuaYHYanjun Hua
- Giacomo SaviniGSGiacomo Savini
- Hao ChenHCHao Chen
- Fangjun BaoFBFangjun Bao
- Shishi LinSLShishi Lin
- Weicong LuWLWeicong Lu
- Jinhai HuangJHJinhai Huang
- 1 October 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Cornea
- Vol. 34 (10) , 1266-1271
Abstract
To assess the repeatability and reproducibility of corneal power values obtained by a Fourier-domain optical coherence tomography (FD-OCT) system (RTVue) and to compare them with the values obtained by a Scheimpflug camera system (Pentacam HR) and by automated keratometry (IOL Master). Thirty-two eyes from 32 healthy subjects were included in this prospective study. Two experienced observers measured each eye 3 consecutive times with the Pentacam, IOLMaster, and RTVue centered on either the pupil or corneal vertex. The conventional keratometry equivalent (CKE) and anterior (Ka), posterior (Kp), and net (Kn) corneal power values were determined. The corneal power values obtained by the RTVue showed high repeatability (all intraclass correlation coefficient >0.96) and reproducibility (coefficient of variation <1.0%). Pupil-centered FD-OCT performed slightly better than corneal vertex-centered FD-OCT. Mean corneal values had higher reproducibly than any of the individual values. CKE, Ka, Kp, and Kn obtained by FD-OCT were 0.62 to 0.68 diopters (D), 0.70 to 0.76 D, 0.11 to 0.13 D, and 0.93 to 0.94 D higher than those obtained by the Pentacam HR, respectively. CKE and Ka obtained with the RTVue were also 0.60 to 0.74 D higher than those obtained with the IOLMaster, respectively. The corneal power measurements obtained by the RTVue FD-OCT system showed high repeatability and reproducibility. Measurements obtained by FD-OCT with pupil centration were more reproducible than those obtained by FD-OCT with corneal vertex centration. We recommend that pupil-centered FD-OCT be used in clinical applications. Neither RTVue versus Pentacam HR nor RTVue versus IOLMaster can be used interchangeably.Keywords
This publication has 21 references indexed in Scilit:
- Ray-tracing intraocular lens power calculation using anterior segment optical coherence tomography measurementsJournal of Cataract & Refractive Surgery, 2012
- Effects of error in radius of curvature on the corneal power measurement before and after laser refractive surgery for myopiaOphthalmic and Physiological Optics, 2012
- Intraocular lens power calculation after previous myopic laser vision correction based on corneal power measured by Fourier-domain optical coherence tomographyJournal of Cataract & Refractive Surgery, 2012
- Estimation of corneal power after myopic laser refractive surgery: Comparison of methods against back-calculated corneal powerJournal of Cataract & Refractive Surgery, 2011
- Corneal power measurement with Fourier-domain optical coherence tomographyJournal of Cataract & Refractive Surgery, 2010
- Accuracy of Scheimpflug corneal power measurements for intraocular lens power calculationJournal of Cataract & Refractive Surgery, 2009
- Validity of the keratometric index: Evaluation by the Pentacam rotating Scheimpflug cameraJournal of Cataract & Refractive Surgery, 2008
- Intraocular lens power calculation after radial keratotomy: Estimating the refractive corneal powerJournal of Cataract & Refractive Surgery, 2007
- Intraocular lens power calculation after corneal refractive surgeryCurrent Opinion in Ophthalmology, 2004
- Correlation between refractive and measured corneal power changes after myopic excimer laser photorefractive surgeryJournal of Cataract & Refractive Surgery, 2002