Case Report: We report a fifty-year-old male with paroxysmal atrial fibrillation presenting with EHRA III symptoms who underwent pulmonary vein isolation (PVI). The PVI was planned with a 28mm cryoballon under deep sedation. After uncomplicated transseptal puncture, angiography of the left atrium (LA) was done showing four independent pulmonary veins (PV). Heparin was given to maintain an activated clotting time of >300s. Following successful left PV isolation, the circular diagnostic catheter (CDC) was advanced in the right inferior PV (RIPV) and one successful freeze was delivered. After thawing of the cryoballon, it turned out not to be possible to remove the CDC from the RIPV. The fluoroscopy showed that the CDC was stuck in a small branch of the RIPV. While pulling the CDC the patient started to cough. The surrounding lung tissues seemed to be tightened and dragged together with the CDC. While turning the CDC, the shaft of the CDC broke close to the holder of the steerable sheath. With use of the remaining shaft, we further tried to remove the CDC from the PV, however, the CDC was again completely torn from the base of the circular part in the branch of RIPV. We then removed the proximal shaft and isolated the superior right PV with one freeze using a newly unpacked CDC. Final LA angiography demonstrated neither obstructions of major PVs nor exit of contrast agent out of LA, although the circular part of the CDC remained in the small branch of RIPV. After the procedure the patient showed no complaints such as cough, dyspnea or other symptoms. A pericardial effusion was also excluded. Discussion: Due to the tying in a small branch of the RIPV the CDC was stuck and we could not recover the catheter anymore. No practical way to resolve this situation has been described yet. Further catheter pulling was not an option due to the risk of a fatal bleeding because the surrounding lung tissue was tied. Another option could be an open-heart surgery. As the remaining part of the catheter did not move at all, we decided to leave the remaining circular part after a consultation with heart surgery unit. After a follow-up of one month the patient showed neither AF event nor clinical symptoms such as cough, dyspnea. We here reported a break of the CDC shaft during PVI with cryoballoon, which has been not yet reported.