An improved physics-based approach for unfolding megavoltage bremsstrahlung spectra using transmission analysis
Authors:
- E. S. M. AliEAE. S. M. Ali
- D. W. O. RogersDRD. W. O. Rogers
- 1 March 2012
- journal article
- Published by Wiley in Medical Physics
- Vol. 39 (3) , 1663-1675
Abstract
To develop a physics-based approach to improve the accuracy and robustness of the ill-conditioned problem of unfolding megavoltage bremsstrahlung spectra from transmission data. Spectra are specified using a rigorously-benchmarked functional form. Since ion chambers are the typical detector used in transmission measurements, the energy response of a Farmer chamber is calculated using the EGSnrc Monte Carlo code, and the effect of approximating the energy response on the accuracy of the unfolded spectra is studied. A proposal is introduced to enhance spectral sensitivity by combining transmission data measured with multiple detectors of different energy response and by combining data from multiple attenuating materials. Monte Carlo methods are developed to correct for nonideal exponential attenuation (e.g., scatter effects and secondary attenuation). The performance of the proposed methods is evaluated for a diverse set of validated clinical spectra (3.5-25 MV) using analytical transmission data with simulated experimental noise. The approximations commonly used in previous studies for the ion-chamber energy response lead to significant errors in the unfolded spectra. Of the configurations studied, the one with best spectral sensitivity is to measure four full transmission curves using separate low-Z and high-Z attenuators in conjunction with two detectors of different energy response (the authors propose a Farmer-type ion chamber, once with a low-Z, and once with a high-Z buildup cap material), then to feed the data simultaneously to the unfolding algorithm. Deviations from ideal exponential attenuation are as much as 1.5% for the smallest transmission signals, and the proposed methods properly correct for those deviations. The transmission data with enhanced spectral sensitivity, combined with the accurate and flexible spectral functional form, lead to robust unfolding without requiring a priori knowledge of the spectrum. Compared with the commonly-used methods, the accuracy is improved for the unfolded spectra and for the unfolded mean incident electron kinetic energy by at least factors of three and four, respectively. With simulated experimental noise and a lowest transmission of 1%, the unfolded energy fluence spectra agree with the original spectra with a normalized root-mean-square deviation, %Δ(ψ), of 2.3%. The unfolded mean incident electron kinetic energies agree, on average, with the original values within 1.4%. A lowest transmission of only 10% still allows unfolding with %Δ(ψ) of 3.3%. In the presence of realistic experimental noise, the proposed approach significantly improves the accuracy and robustness of the spectral unfolding problem for all therapy and MV imaging beams of clinical interest.Keywords
This publication has 33 references indexed in Scilit:
- Influence of ion chamber response on in-air profile measurements in megavoltage photon beamsMedical Physics, 2005
- Enhanced spectral discrimination through the exploitation of interface effects in photon dose dataMedical Physics, 2004
- Reconstruction of 12 MV bremsstrahlung spectra from measured transmission data by direct resolution of the numeric system AF=TMedical Physics, 1995
- Simulation of x-ray spectral reconstruction from transmission data by direct resolution of the numeric system AF =TMedical Physics, 1993
- Dose perturbations at interfaces in photon beamsMedical Physics, 1987
- Estimated spectrum of a 4-MV therapeutic beamMedical Physics, 1986
- Application of a Laplace transform pair model for high-energy x-ray spectral reconstructionMedical Physics, 1985
- Reconstruction of 4‐MV bremsstrahlung spectra from measured transmission dataMedical Physics, 1983
- Simulation studies of 4‐MV x‐ray spectral reconstruction by numerical analysis of transmission dataMedical Physics, 1982
- Spectral characterization of 4 MV Bremsstrahlung by attenuation analysisMedical Physics, 1981