Sampling strategies and integrated reconstruction for reducing distortion and boundary slice aliasing in high‐resolution 3D diffusion MRI
Open Access
- 15 June 2023
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 90 (4) , 1484-1501
Abstract
Purpose: To develop a new method for high‐fidelity, high‐resolution 3D multi‐slab diffusion MRI with minimal distortion and boundary slice aliasing.Methods: Our method modifies 3D multi‐slab imaging to integrate blip‐reversed acquisitions for distortion correction and oversampling in the slice direction (kz) for reducing boundary slice aliasing. Our aim is to achieve robust acceleration to keep the scan time the same as conventional 3D multi‐slab acquisitions, in which data are acquired with a single direction of blip traversal and without kz‐oversampling. We employ a two‐stage reconstruction. In the first stage, the blip‐up/down images are respectively reconstructed and analyzed to produce a field map for each diffusion direction. In the second stage, the blip‐reversed data and the field map are incorporated into a joint reconstruction to produce images that are corrected for distortion and boundary slice aliasing.Results: We conducted experiments at 7T in six healthy subjects. Stage 1 reconstruction produces images from highly under‐sampled data (R = 7.2) with sufficient quality to provide accurate field map estimation. Stage 2 joint reconstruction substantially reduces distortion artifacts with comparable quality to fully‐sampled blip‐reversed results (2.4× scan time). Whole‐brain in‐vivo results acquired at 1.22 mm and 1.05 mm isotropic resolutions demonstrate improved anatomical fidelity compared to conventional 3D multi‐slab imaging. Data demonstrate good reliability and reproducibility of the proposed method over multiple subjects.Conclusion: The proposed acquisition and reconstruction framework provide major reductions in distortion and boundary slice aliasing for 3D multi‐slab diffusion MRI without increasing the scan time, which can potentially produce high‐quality, high‐resolution diffusion MRI.Keywords
All Related Versions
Funding Information
- Royal Academy of Engineering (RF∖201819∖18∖92)
- Wellcome Trust (203139/Z/16/Z, WT202788/Z/16/A)
This publication has 47 references indexed in Scilit:
- Coil compression for accelerated imaging with Cartesian samplingMagnetic Resonance in Medicine, 2012
- Blipped‐controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g‐factor penaltyMagnetic Resonance in Medicine, 2011
- Diffusion imaging of whole, post-mortem human brains on a clinical MRI scannerNeuroImage, 2011
- SPIRiT: Iterative self‐consistent parallel imaging reconstruction from arbitrary k‐spaceMagnetic Resonance in Medicine, 2010
- Efficient correction of inhomogeneous static magnetic field-induced distortion in Echo Planar ImagingNeuroImage, 2010
- High efficiency, low distortion 3D diffusion tensor imaging with variable density spiral fast spin echoes (3D DW VDS RARE)NeuroImage, 2010
- Comprehensive quantification of signal‐to‐noise ratio and g‐factor for image‐based and k‐space‐based parallel imaging reconstructionsMagnetic Resonance in Medicine, 2008
- Probabilistic diffusion tractography with multiple fibre orientations: What can we gain?NeuroImage, 2007
- Diffusional kurtosis imaging: The quantification of non‐gaussian water diffusion by means of magnetic resonance imagingMagnetic Resonance in Medicine, 2005
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002