Comparison of Single‐Coil Versus Dual‐Coil Implantable Cardioverter Defibrillator Devices: A Systematic Review and Meta‐Analysis of Efficacy and Extraction‐Related Outcomes
Authors:
- Muhammad Fawad TahirMTMuhammad Fawad Tahir
- Adeena MaryyumAMAdeena Maryyum
- Zainab MubbashirZMZainab Mubbashir
- Abdul Moiz KhanAKAbdul Moiz Khan
- Syed Irtiza ImamSISyed Irtiza Imam
- Fatima MustafaFMFatima Mustafa
- Syeda Zahra HasanSHSyeda Zahra Hasan
- Umer ShoaibUSUmer Shoaib
- Osama SaeedOSOsama Saeed
- Manisha PurushothamMPManisha Purushotham
- Maimoona KhanMKMaimoona Khan
Open Access
- 5 February 2025
- journal article
- review article
- Published by Wiley in Clinical Cardiology
- Vol. 48 (2) , e70083
Abstract
Background: Implantable cardioverter defibrillators (ICD) are battery‐operated devices used to manage irregular heart rhythms and deliver therapeutic shocks to the heart. This updated systematic review and meta‐analysis compares the efficacy and extraction‐related outcomes of single‐coil versus dual‐coil ICDs in view of conflicting data.Methods: Several databases, including PubMed, Cochrane Library, and Google Scholar, were comprehensively explored dating from inception to April 1, 2024. Data were compared using odds ratio (OR), hazard ratio (HR), and mean differences (MD). A value of p < 0.05 indicated statistical significance.Results: Ultimately, 28 studies were included in this quantitative synthesis. Defibrillation threshold (DFT) indicated statistical superiority in the dual‐coil cohort (MD: 0.58; 95% confidence interval [CI]: 0.07–1.09; p = 0.03). In addition, all‐cause mortality was significantly elevated in the dual‐coil cohort (HR: 0.91; 95% CI: 0.87–0.97; p = 0.001). Furthermore, implant time was significantly lower in the single‐coil group (MD: −7.44; 95% CI: −13.44 to −1.43; p = 0.02). Other outcomes, including first shock efficacy, cardiac mortality, post‐extraction major complications, post‐extraction procedural success, and post‐extraction mortality, did not demonstrate any significant statistical differences.Conclusion: In conclusion, despite the desirable safety profile of single‐coil ICDs, the use of dual‐coil ICDs continues to hold merit due to their superior efficacy and advanced sensing capabilities, especially in complex cases. In addition, the perceived risk of a greater adverse profile in dual‐coil lead extraction can be refuted by preliminary aggregate results generated within this meta‐analysis. However, further robust studies are warranted before arriving at a valid conclusion.Keywords
This publication has 46 references indexed in Scilit:
- A Randomized Prospective Study of Single Coil Versus Dual Coil Defibrillation in Patients with Ventricular Arrhythmias Undergoing Implantable Cardioverter Defibrillator TherapyPacing and Clinical Electrophysiology, 2003
- Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection FractionNew England Journal of Medicine, 2002
- Dual-coil vs single-coil active pectoral implantable defibrillator lead systems: defibrillation energy requirements and probability of defibrillation success at multiples of the defibrillation energy requirementsEP Europace, 2001
- Comparison of defibrillation thresholds using monodirectional electrical vector versus bidirectional electrical vector.2001
- Safety and efficacy of low-energy cardioversion of 500 patients using two different techniquesEP Europace, 2001
- Two-Coil Versus Single-Coil Transvenous Cardioverter Defibrillator Systems: Comparative DataPacing and Clinical Electrophysiology, 2000
- Optimization of Transvenous Coil Position for Active Can Defibrillation ThresholdsJournal of Cardiovascular Electrophysiology, 2000
- Comparison of Single- and Dual-Coil Active Pectoral Defibrillation Lead SystemsJACC, 1998
- Lead System Optimization for Transvenous DefibrillationThe American Journal of Cardiology, 1997
- Prospective, randomized comparison in humans of a unipolar defibrillation system with that using an additional superior vena cava electrode.Circulation, 1994