Wear debris released by hip prosthesis analysed by microcomputed tomography
Authors:
- 29 October 2020
- journal article
- research article
- Published by Wiley in Journal of Microscopy
- Vol. 282 (1) , 13-20
Abstract
Summary: Total hip arthroplasty uses commercial devices that combine different types of biomaterials. Among them, metals, ceramics and metal oxides can be used either in the prosthesis itself or in the cement used to anchor them in the bone. Over time, all of these materials can wear out and release particles that accumulate in the periprosthetic tissues or can migrate away. We used histology blocks from 15 patients (5 titanium metallosis, 5 alumina prostheses, 5 with altered methacrylic cement) to perform a microCT study and compare it with conventional histology data. An EDS‐SEM analysis was done to characterise the atomic nature of the materials involved. A morphometric analysis was also performed in 3D to count the particles and assess their density and size. The metallic particles appeared to be the largest and the ceramic particles the finest. However, microCT could not reveal the wear particles of radiolucent biomaterials such as polyethylene and the very fine zirconia particles from cement fragmentation. MicroCT analysis can reveal the extent of the accumulation of these debris in the periprosthetic tissues.Layout Description: Hip prostheses progressively degrade in the body by releasing wear debris. They accumulate in the periprosthetic tissues. Microcomputed tomography was used to image three types of radio‐opaque wear debris: metal, ceramic and zirconia used in the bone cements.Keywords
This publication has 28 references indexed in Scilit:
- The Role of Micro-CT in 3D Histology ImagingPathobiology, 2016
- Release of zirconia nanoparticles at the metal stem–bone cement interface in implant loosening of total hip replacementsActa Biomaterialia, 2016
- Total arthroplasty of the hip by fritted alumina prosthesis. Experimental study and 1st clinical applicationsOrthopaedics & Traumatology: Surgery & Research, 2014
- Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstreamInternational Orthopaedics, 2013
- Metallosis After Contemporary Metal-on-Metal Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2006
- Prevention of Wear and OsteolysisClinical Orthopaedics and Related Research, 2001
- Cytotoxicities of oxides, phosphates and sulphides of metalsBiomaterials, 1992
- Fragmentation of methylmethacrylate: A cause of late failure of total hip replacementArchives of orthopaedic and trauma surgery, 1989
- Wear analysis of retrieved alumina heads and sockets of hip prosthesesJournal of Biomedical Materials Research, 1989
- Surgery of the Hip-jointBMJ, 1960