Monte Carlo calculation of detector perturbation and quality correction factors in a 1.5 T magnetic resonance guided radiation therapy small photon beams
Open Access
- 9 November 2021
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 66 (22) , 225004
Abstract
Objective. With future advances in magnetic resonance imaging-guided radiation therapy, small photon beams are expected to be included regularly in clinical treatments. This study provides physical insights on detector dose-response to multiple megavoltage photon beam sizes coupled to magnetic fields and determines optimal orientations for measurements. Approach. Monte Carlo simulations determine small-cavity detector (solid-state: PTW60012 and PTW60019, ionization chambers: PTW31010, PTW31021, and PTW31022) dose-responses in water to an Elekta Unity 7 MV FFF photon beam. Investigations are performed for field widths between 0.25 and 10 cm in four detector axis orientations with respect to the 1.5 T magnetic field and the photon beam. The magnetic field effect on the overall perturbation factor (P MC) accounting for the extracameral components, atomic composition, and density is quantified in each orientation. The density (P ρ ) and volume averaging (P vol) perturbation factors and quality correction factors ( k Q B , Q f B , f ) accounting for the magnetic field are also calculated in each orientation. Main results. Results show that P vol remains the most significant perturbation both with and without magnetic fields. In most cases, the magnetic field effect on P vol is 1% or less. The magnetic field effect on P ρ is more significant on ionization chambers than on solid-state detectors. This effect increases up to 1.564 ± 0.001 with decreasing field size for chambers. On the contrary, the magnetic field effect on the extracameral perturbation factor is higher on solid-state detectors than on ionization chambers. For chambers, the magnetic field effect on P MC is only significant for field widths Significance. Solid-state detectors’ dose-response is strongly affected by the magnetic field in all orientations. The magnetic field impact on ionization chamber response increases with decreasing field size. In general, ionization chambers yield k Q B , Q f B , f closer to unity, especially in orientations where the chamber axis is parallel to the magnetic field.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (NSERC RGPIN/04178-2015)
- UK government’s Department for Business, Energy and Industrial Strategy and the EMPIR programme (Grant 19NRM01)
- Co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme
This publication has 39 references indexed in Scilit:
- Towards reference dosimetry for the MR-linac: magnetic field correction of the ionization chamber readingPhysics in Medicine & Biology, 2013
- Using cavity theory to describe the dependence on detector density of dosimeter response in non-equilibrium small fieldsPhysics in Medicine & Biology, 2013
- Characterizing the influence of detector density on dosimeter response in non-equilibrium small photon fieldsPhysics in Medicine & Biology, 2012
- Calculation of for several small detectors and for two linear accelerators using Monte Carlo simulationsMedical Physics, 2011
- Investigation of systematic uncertainties in Monte Carlo-calculated beam quality correction factorsPhysics in Medicine & Biology, 2010
- Ionization chamber gradient effects in nonstandard beam configurationsMedical Physics, 2009
- Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamberPhysics in Medicine & Biology, 2009
- The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambersPhysics in Medicine & Biology, 2009
- Efficiency improvements for ion chamber calculations in high energy photon beamsMedical Physics, 2008
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005