Measuring the Forces of Middle Ear Surgery; Evaluating a Novel Force-Detection Instrument
Authors:
- Michael SheedyMSMichael Sheedy
- Peter RossPRPeter Ross
- Grant WylieGWGrant Wylie
- Philip BirdPBPhilip Bird
- 1 February 2014
- journal article
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 35 (2) , e77-e83
Abstract
The forces applied to the middle ear structures during surgery are measurable. Surgical forces applied to the middle ear are often cited as a cause of postoperative sensorineural hearing loss; however, no literature exists on how much total force is applied during common middle ear procedures. Using our novel middle ear surgical force sensor, we took measurements from 6 temporal bones of common middle ear manipulations. Temporal bones of 3 males and 3 females yielded 72 measurements for an average peak applied force and 68 measurements for total applied force across 3 levels of operator experience. Statistically significant differences were seen between different levels of operator and how much force was applied, with senior operators using less force. We show it is possible to measure the applied forces in middle ear surgery and that some of these forces are very large. Larger forces are seen in destructive procedures and in procedures performed by inexperienced operators. This project is an important first step in investigating the association between surgically applied middle ear forces and postoperative sensorineural hearing loss.Keywords
This publication has 20 references indexed in Scilit:
- Surgical Instrument Force Exerted during Endoscopic Sinus Surgery: Differences within the Ethmoid Sinus ComplexAmerican Journal of Rhinology & Allergy, 2010
- A sub-millimetric, 0.25 mN resolution fully integrated fiber-optic force-sensing tool for retinal microsurgeryInternational Journal of Computer Assisted Radiology and Surgery, 2009
- A miniature microsurgical instrument tip force sensor for enhanced force feedback during robot-assisted manipulationIEEE Transactions on Robotics and Automation, 2003
- Drilling in Ear Surgery: A Comparison of Pre- and Postoperative Bone-conduction Thresholds in both the Conventional and Extended High-frequency RangesScandinavian Audiology, 1996
- Drill‐generated sensorineural hearing loss following mastoid surgeryThe Laryngoscope, 1992
- Hearing aids poststapedectomy: Incidence and timingThe Laryngoscope, 1986
- Ossicular Manipulation in Chronic Ear Surgery: An Electrocochleographic StudyJAMA Otolaryngology–Head & Neck Surgery, 1980
- Middle-Ear Characteristics of Anesthetized CatsThe Journal of the Acoustical Society of America, 1967
- Stapedectomy for Otosclerosis: Some Causes of FailureJAMA Otolaryngology–Head & Neck Surgery, 1967
- Acoustic trauma of the cochlea from ear surgeryThe Laryngoscope, 1960