Development of an algorithm to convert mammographic images to appear as if acquired with different technique factors
- 8 June 2022
- journal article
- research article
- Published by SPIE-Intl Soc Optical Eng in Journal of Medical Imaging
- Vol. 9 (03) , 033504-19
Abstract
A key step in the setting up of a mammography system is optimization, i.e., to ensure that the quality of the images is sufficient for an accurate diagnosis for as low a dose as reasonably practicable. There are a number of choices of factors to be made in optimization such as the tube voltage, anode, and filter materials, the use of an anti-scatter grid, and the air kerma at the detector. The technique factors will affect the signal level and the difference in signal between adipose and glandular tissues and any lesions. Noise arising from the detector can affect the visibility of low contrast objects in an image. The dominant detector noise is quantum noise, which is dependent on the signal level and x-ray spectra incident on the detector.1 A method used for optimization of the tube voltage and anode/filter combination for mammography is to image a block of polymethyl methacrylate (PMMA) with a low-contrast object, such as 0.2-mm thick aluminum, over a range of technique factors and use a figure of merit based on the contrast-to-noise ratio (CNR) and breast dose.2,3 Such measurements have indicated that digital systems should use anode/filter combinations that produce higher beam energies than those traditionally used in screen/film mammography (e.g., molybdenum/molybdenum or molybdenum/rhodium).3Keywords
This publication has 32 references indexed in Scilit:
- X‐ray scatter correction in breast tomosynthesis with a precomputed scatter map libraryMedical Physics, 2014
- Effect of image quality on calcification detection in digital mammographyMedical Physics, 2012
- Conversion of mammographic images to appear with the noise and sharpness characteristics of a different detector and x‐ray systemMedical Physics, 2012
- Visibility of microcalcification clusters and masses in breast tomosynthesis image volumes and digital mammography: A 4AFC human observer studyMedical Physics, 2012
- Detective quantum efficiency measured as a function of energy for two full-field digital mammography systemsPhysics in Medicine & Biology, 2009
- The effect of scatter and glare on image quality in contrast‐enhanced breast imaging using an full‐field flat panel detectorMedical Physics, 2009
- Evaluation of clinical image processing algorithms used in digital mammographyMedical Physics, 2009
- Use of the European Protocol to Optimise a Digital Mammography SystemLecture Notes in Computer Science, 2006
- A framework for optimising the radiographic technique in digital X-ray imagingRadiation Protection Dosimetry, 2005
- Method of simulating dose reduction for digital radiographic systemsRadiation Protection Dosimetry, 2005