Abstract
A key step in the setting up of a mammography system is optimization, i.e., to ensure that the quality of the images is sufficient for an accurate diagnosis for as low a dose as reasonably practicable. There are a number of choices of factors to be made in optimization such as the tube voltage, anode, and filter materials, the use of an anti-scatter grid, and the air kerma at the detector. The technique factors will affect the signal level and the difference in signal between adipose and glandular tissues and any lesions. Noise arising from the detector can affect the visibility of low contrast objects in an image. The dominant detector noise is quantum noise, which is dependent on the signal level and x-ray spectra incident on the detector.1 A method used for optimization of the tube voltage and anode/filter combination for mammography is to image a block of polymethyl methacrylate (PMMA) with a low-contrast object, such as 0.2-mm thick aluminum, over a range of technique factors and use a figure of merit based on the contrast-to-noise ratio (CNR) and breast dose.2,3 Such measurements have indicated that digital systems should use anode/filter combinations that produce higher beam energies than those traditionally used in screen/film mammography (e.g., molybdenum/molybdenum or molybdenum/rhodium).3